Programmer's Manual

Table Of Contents
92 English Programmer Model 2740
The programmer is not certified for use in the presence of a flammable anesthetic mixture with air or
with oxygen or nitrous oxide. The consequences of using the programmer near flammable
atmospheres are unknown.
Operate the system at the recommended temperature range of 5°C to 30°C (41°F to 86°F). Store it at
a temperature of -20°C to 60°C (-4°F to 140°F).
Certain environments may contain particles or debris that can adhere to the tablet LCD screen. To
avoid damaging the display, clean it frequently by diluting a household, water-based glass cleaner to a
50/50 ratio with water, and spray a small amount on a clean, soft cloth and gently wipe the screen.
Always shut down and unplug the programmer before cleaning. Do not spray cleaner directly on the
tablet.
Electromagnetic disturbances from the programmer may interfere with other equipment or the
programmer could be interfered with by other equipment, including portable and mobile Radio
Frequency (RF) communications equipment. If interference occurs, relocate the equipment.
Do not bring the programmer into Zone 4 (magnet room), as defined by the American College of
Radiology. The programmer is MR Unsafe.
Electromagnetic interference could be caused by unseen sources such as radio frequency identification
(RFID) devices or wireless charging (Qi). If interference is suspected, relocate the equipment and use
the Test Connection feature to evaluate connection performance.
Do not open or attempt to service this product. Opening or servicing the system components can result
in electric shock.
Restrict access to the Model 2740 to authorized users as the tablet does not provide user
authentication. Tablet Wi-Fi and general-purpose Bluetooth functionality are disabled, therefore
network security controls such as firewalls are not required. Tablet operation system (Microsoft
Windows) is protected using write filtering, therefore, anti-virus software is not required nor
supported. Return system components to Inspire Medical Systems for secure decommissioning. Notify
Inspire Medical Systems of any suspected security event.
Disposal
Do not dispose of the system components if they are no longer being used or if they become
inoperable. They must be returned to Inspire Medical Systems.
Dispose of the tablet (containing a lithium-ion battery pack) only at approved disposal sites. To locate
an appropriate site, contact the solid waste disposal officials where you live, or look for a rechargeable
battery recycling Web site that lists disposal locations.
Do not dispose of tablets (containing a lithium-ion battery pack) in a fire. They may explode. Dispose
of used batteries according to the manufacturer’s instructions. The label on the battery lists the
manufacturer’s name.