Programmer's Manual

Table Of Contents
Programmer Model 2740 English 93
Power Cord and Other Cables
Do not touch mains connected parts (power cords) and the patient simultaneously because of a risk of
electrical shock to the patient and user.
Power cord sets used in other countries must meet the requirements of that country. Use the
appropriate power cord for your locale. For information about power cord set requirements, contact
Inspire Medical Systems.
Do not plug the power cord into an extension cord or multiple portable socket outlet (MPSO). The
device has not been tested for safety or electromagnetic emissions in this configuration, and proper
performance cannot be guaranteed. Plug the power cord directly into an electrical outlet.
When using the power cord, make sure to position it around objects so it will not be cut or punctured.
Make sure the connection where the power cord connects to the mains power is easily accessible and
can be easily disconnected by the user.
Tablet Battery Pack
The tablet contains an internal lithium-ion (Li-ion) battery pack. Do not attempt to remove or replace
the battery pack.
Always store the programmer with the battery charged and attached to the tablet.
If the programmer is stored for more than 4 weeks without use, connect it to power until it is fully
charged. The tablet does not need to be turned on to charge.
Do not use or leave the tablet and its battery pack near a heat source. Heat can melt the insulation and
damage other safety features, possibly causing it to leak, overheat, emit smoke, burst and/or ignite.