Programmer Model 2740 English 97
Device Specifications
Additional equipment connected to medical electrical equipment must comply with the respective IEC or ISO
standards (e.g., IEC 60950 for data processing equipment). Furthermore, all configurations shall comply with the
requirements for medical electrical systems (see IEC 60601-1-1 or clause 16 of the third edition of IEC 60601-1,
respectively).
Anyone connecting additional equipment to medical electrical equipment configures a medical system and is
therefore responsible that the system complies with the requirements for medical electrical systems.
Local laws take priority over the requirements mentioned above. Contact Inspire Medical Systems with any
questions.
Tablet
Table 8-2. Tablet Specifications
a
a
All measurements are approximate
Description Specifications
Temperature with power on
Temperature: 5°C–30°C (41°F–86°F)
Humidity: 20–80%
Air pressure: 700–1013 hPa (10.2–14.7 psi)
Temperature with power off for
storage or transportation
Temperature: -20°C–60°C (-4°F–140°F)
Humidity: RH 10–90% RH
Air pressure: 700–1013 hPa (10.2–14.7 psi)
Tablet power supply Sinpro
M/N: HPU3
*-105
(UL agency approval under
UL 60601-1)
Input: 100–240 VAC, 47–63 Hz, 0.6-0.4A.
Output: 12V, 2.5A.
Ingress protection IP65
Mode of operation
(continuous or non-continuous)
Continuous