Programmer Model 2740 English 101
Electromagnetic Compatibility Declarations
The Inspire Programmer (Model 2740) utilizes RF communications between the tablet and programmer cable. RF
communications utilize the 2.4 to 2.485GHz ISM band and frequency shift keying modulation. Frequency-hopping
spread spectrum is utilized to avoid interference with other devices. Effective radiated power from RF
communications are less than 10mW.
Additional non-standardized testing was performed to demonstrate that the Bluetooth wireless communications
of the programmer can coexist with potential interfering devices like a Wi-Fi router. With the source of
interference located 1 meter (3 feet) from the programmer tablet or programmer cable, the programmer’s
Bluetooth suffered no decrease in performance.
Table 8-4. Electromagnetic Emissions
The Inspire Programmer (Model 2740) is intended for use in the electromagnetic environment
specified below. The customer or the user of the programmer should ensure that it is used in
such an environment.
Emissions Test Compliance Electromagnetic Environment Guidance
Radio frequency
(RF) emissions
CISPR 11
Group 1 The programmer uses RF energy for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions CISPR
11
Class A The programmer is suitable for use in
professional healthcare facility environments.
Note: The emissions characteristics of this
equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used
in a residential environment (for which CISPR 11
class B is normally required) this equipment
might not offer adequate protection to radio
frequency communication services. The user
might need to take mitigation measures, such as
relocating or re-orienting the equipment.
Warning: This system is intended for use by
health care professionals only.
Harmonic
emissions
EN 61000-3-2
Class A
Voltage
fluctuations/flicker
emissions EN
61000-3-3
Complies