Table 8-5. Electromagnetic Immunity
The Inspire Programmer (Model 2740) is intended for use in the electromagnetic environment specified below.
The customer or the user of the programmer should ensure that it is used in such an environment.
Immunity Test EN/IEC 60601 Test
Level
Compliance Level Electromagnetic Environment Guidance
Electrostatic
discharge (ESD):
EN/IEC
61000-4-2
+/-8 kV contact
+/-15 kV air
+/-8 kV contact
+/-15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrostatic discharge may result in a
temporary loss of function, requiring the
user to restart the programmer.
Electrical fast
transient/burst:
EN/IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
+/-2kV for input AC
power port
+/-1 kV for
signal I/O port
100 kHz repetition
frequency
±2 kV for power
supply lines
(no input/output
lines)
+/-2kV for input AC
power port
+/-1 kV for
signal I/O port
100 kHz repetition
frequency
Mains power quality should be that of a
typical commercial or hospital
environment.
Electrical fast transients or bursts in mains
power may result in temporary loss of
function.
Surge: EN/IEC 61000-
4-5
+/-0.5 kV, +/-1
kV line-to-line
+/-0.5 kV, +/-1
kV, +/-2 kV
line-to-ground
+/-0.5 kV, +/-1kV
line-to-line
+/-0.5 kV, +/-1kV,
+/-2 kV
line-to-ground
Mains power quality should be that of a
typical commercial or hospital
environment.