Table 8-6. Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Inspire Programmer (Model 2740) is intended for use in the electromagnetic environment specified
below. The customer or the user of the programmer should ensure that it is used in such an environment.
Immunity Test EN/IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the programmer, including cables, than
the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey
a
, should be less than the compliance
level in each frequency range
b
.
Conducted RF
EN/IEC 61000-
4-6
6 Vrms in ISM
bands between
0.15 MHz and
80 MHz
80% AM at
1 kHz
6 Vrms in ISM
bands between
0.15 MHz and
80 MHz
80% AM at
1 kHz
Radiated RF EN/
IEC
61000-4-3
3 V/m
80 MHz - 2.7
GHz
80% AM at
1 kHz
3 V/m
80 MHz - 2.7
GHz
80% AM at
1 kHz
Power
frequency
magnetic field:
EN/IEC 610000-
4-8
30 A/m 30 A/m
d 1.2 P=
d 1.2 P=
d 2.3 P=