Programmer's Manual

Table Of Contents
24 English Programmer Model 2740
Setting Data Controls
Select the Settings button from the Start screen to adjust the following:
USB Export — Select this check box to enable or disable the option to transfer session reports to the
Inspire USB flash drive.
Patient Information Storage Select this check box to enable or disable the storage/editing of patient
name and ID.
Patient Information Export Select this check box to enable or disable the exporting of patient details.
Select the Save button to save the new data control settings.
Setting Language
1. Select the Settings button from the Start screen.
2. Select the Language menu and choose one of the languages.
3. Select the Save button to save the new language.
Setting Number Format
1. Select the Settings button from the Start screen.
2. Select the Number Format drop-down menu and choose one of the formats that displays below.
3. Select the Save button to save the new number format.
Setting Date and Time
1. Select the Settings button from the Start screen.
2. Select the arrow buttons to adjust the day, month and year.
3. Select the arrow buttons to adjust the 24-hour clock.
4. Select the Save button to save the new date and time.
Note: The programmer does not automatically adjust for daylight savings time. Manually update the date
and time when local time changes occur.
Positioning Programmer Head and Connecting to Generator
The programmer head must be properly positioned over the generator to establish the signal strength required
to connect to the generator.
Confirming Connection
The programmer cable status line in the lower-left corner of the screen will indicate that a communication
link between the tablet and the programmer cable is established. See “Programmer Cable Status” on
page 22“ for more information.
A fully lit status light displays on the programmer cable controller when the tablet and programmer cable
establish a connection.