Programmer's Manual

Table Of Contents
Programmer Model 2740 English 27
A large communication icon displays on all Communication Screens (Figure 3-8). The icon corresponds to the
programmer head status light to indicate communication success or failure.
Figure 3-8. Communication Screen
Table 3-4 describes common problems associated with establishing a wireless connection between the tablet,
programmer cable, and generator along with their solutions.
Table 3-3. Communication Screens
Message Event Action
Connecting to generator Displays when the
Connect to Generator
button has been selected.
Wait for connection to be established.
Configuring generator Displays when the
generator is being updated
for the start of a session.
Wait for connection to be established.
Generator Not Found or
Electrical Interference
Displays when the
programmer head is not
correctly positioned over
the generator.
Reposition programmer head over generator
or remove source of interference. After 30
seconds of communication failure, the Exit
Session button is available to end the current
session.