Programmer's Manual

Table Of Contents
28 English Programmer Model 2740
Using Demonstration Mode
Use Demonstration Mode to practice using the programmer with a simulated generator. The programmer cable
is not needed for Demo Mode.
1. Ensure that the tablet is on.
2. Select the Demo button from the Start screen.
Note: Demonstration Mode is displayed in the programmer cable status line located in the lower-left
corner of the tablet screen.
Table 3-4. Communications Troubleshooting
Problem Action
No Connection Between Tablet
and Programmer Cable
Ensure that the tablet is within wireless range. Move tablet closer
to programmer cable if necessary.
Ensure programmer cable is powered on. Check for loose cord
connections and restore power to the programmer cable.
Communication Buttons
Disabled
Connect programmer cable to power.
Ensure tablet is within wireless range to establish connection and
enable all communication buttons.
Generator Not Found Reposition programmer head directly over generator.
Electrical Interference Remove cause of interference and reposition programmer head
directly over generator. Electrical equipment, such as inductive
plethysmography, CRT monitors, and electrical power supplies are
possible causes of interference.