Exiting a Session and Turning Off the Programmer
Allow all communications to complete before a patient leaves. If a patient is allowed to leave during a test or
impedance measurement, the generator may not be restored to therapy settings. Verify that all setting changes
have been made on the Home screen before exiting a session.
Always end sessions properly, according to the following instructions:
1. From any screen, select the Home button to access the Home screen.
2. Select the Exit button to end the session and display the Start screen.
3. Select the Shut Down button to power off the tablet.
Note: The tablet will continue to charge when powered off and should be fully charged before storing. The
charging light will turn off when the tablet is fully charged.
4. To power off the programmer cable and disconnect it from mains power, unplug the power cord from the
electrical outlet.
5. To disconnect the tablet power supply from mains power, unplug the power cord from the electrical
outlet.
Note: The programmer is not designed for use over extended periods of time (>24 hours) without restarting.