Adjust Stimulation Screen
The Adjust Stimulation screen is used to modify and test stimulation settings (Figure 4-11). The setting values
initially displayed on the screen are the values currently in use by the generator. When setting values are changed
and differ from the current generator settings, the fields are highlighted in blue (Figure 4-12). Selecting the
Configure button will configure the generator with the highlighted values and clear the highlighting.
Figure 4-11. Adjust Stimulation Screen, Basic View
Basic Screen View
The basic screen view divides the settings into two sections: stimulation and timing.
Stimulation
Use the arrow buttons to increase or decrease amplitude values.
Select the Enable Patient Control box to allow (check) or disallow (uncheck) patients to adjust
amplitude within a predetermined range.
Use the arrow buttons to set the upper and lower limits of this range.
When Patient Control is enabled, changing the amplitude value will cause the upper and lower
limit values to change as well, tracking the amplitude value.
The values displayed in parentheses next to the upper and lower limits indicate the relative value,
or difference, between the amplitude value and the limit value. The difference will remain constant
as the limit values track changes in the amplitude.
Review flags indicate that the patient has changed the amplitude value, and thus the amplitude values
should be reviewed.