Timing
Use the arrow buttons to increase or decrease the following timing settings.
Start Delay
Therapy Duration
Pause Time
See Table 1-2 for definitions of these terms.
Figure 4-12. Changed Settings on Adjust Stimulation Screen
Programming Buttons
The programming buttons on the Adjust Stimulation screen are disabled until a connection to the programmer
cable is established.
Test Stimulation
The Test Stimulation button requires that the programmer cable is connected. Confirm connection by
checking the programmer cable status in the screen footer (Figure 4-3).
Select the Test Stimulation button to test the displayed settings, which may differ from those currently
configured in the generator. Testing stimulation will not change the current generator settings.
During test stimulation, the generator delivers stimulation for the duration of the maximum
stimulation time.
It is recommended to test stimulation with the therapy off.