44 English Programmer Model 2740
Timing
Use the arrow buttons to increase or decrease the following timing settings.
• Start Delay
• Therapy Duration
• Pause Time
See Table 1-2 for definitions of these terms.
Figure 4-12. Changed Settings on Adjust Stimulation Screen
Programming Buttons
The programming buttons on the Adjust Stimulation screen are disabled until a connection to the programmer
cable is established.
Test Stimulation
• The Test Stimulation button requires that the programmer cable is connected. Confirm connection by
checking the programmer cable status in the screen footer (Figure 4-3).
• Select the Test Stimulation button to test the displayed settings, which may differ from those currently
configured in the generator. Testing stimulation will not change the current generator settings.
• During test stimulation, the generator delivers stimulation for the duration of the maximum
stimulation time.
• It is recommended to test stimulation with the therapy off.