Programmer's Manual
Table Of Contents
- Explanation of Symbols
- How to Use this Guide
- Chapter 1: Therapy Overview
- Chapter 2: Programmer Components
- Chapter 3: Getting Started
- Chapter 4: Screen Descriptions
- Chapter 5: Clinical Programming Sessions
- Chapter 6: Troubleshooting
- Chapter 7: Warnings and Precautions
- Chapter 8: Supplemental Information
- Chapter 9: Inspire Medical Systems, Inc. Limited Warranty
Programmer Model 2740 English 45
Configure
Select the Configure button to configure the generator with all displayed settings that differ from the current
generator settings.
It is important to note that if the Configure button is not pressed before leaving this screen, the setting
changes will not be saved to the generator.
Notes:
• The Configure button requires that the programmer cable is connected. Confirm connection by
checking the programmer cable status in the screen footer (Figure 4-3)
• The Configure button is enabled when changes are made to the settings currently in use by the
generator.
• After selecting the Configure button, the highlighting is cleared because the displayed values now
match the newly configured generator settings.
• When therapy is turned on or setting changes are made with the therapy on, the generator needs
approximately 1–3 minutes to resynchronize with respiration. For this reason, it is recommended to
make changes only after observing therapy performance for 3–5 minutes.
Select a Preset
Use the Select a Preset button to change stimulation settings to one of the following:
• Initial Session Settings — Selects the generator stimulation settings from the start of the current
programming session.
• Current Generator Settings — Selects the stimulation settings currently in use by the generator.
Note: Selecting the Current Generator Settings preset clears all highlighting.
• Default Settings — Selects the default stimulation settings. See “Default Settings” on page 96 for more
information.
Stim History
Select the Stim History button to access stimulation levels and settings stored in the programmer from earlier
in the session or from previous sessions.