Programmer Model 2740 English 49
Adjust Sensing Screen
The Adjust Sensing screen (Figure 4-16) allows for real-time evaluation of the respiratory sensor waveform and
adjustment of sensing settings. Adjustments to sensing settings can modify when stimulation is delivered during
the respiratory cycle.
The setting values initially displayed on the screen are the values currently in use by the generator. When setting
values are changed and differ from the current generator settings, the fields are highlighted in blue. Selecting the
Configure button will configure the generator with the highlighted values and clear the highlighting.
Figure 4-16. Adjust Sensing Screen, Basic View
Basic Screen View
Use the arrow buttons to increase or decrease the following sensing settings.
• Exhalation Sensitivity
• Hard Off Period
Select the Invert Signal box to invert the sensor signal before it is processed.
See Table 1-3 on page 5 for more information about the terminology and settings used on this screen.
Exhalation (Red)
Inhalation (Blue)
Stimulation (Green)
Hard Off Period (Dark)