Programmer Model 2740 English 51
Figure 4-17. Screenshot Screen
Configure
Select the Configure button to configure the generator with all displayed settings that differ from the current
generator settings.
It is important to note that if the Configure button is not pressed before leaving this screen, the setting
changes will not be saved to the generator.
Notes:
The Configure button requires that the programmer cable is connected. Confirm connection by
checking the programmer cable status in the screen footer (Figure 4-3).
The Configure button is enabled when changes are made to the settings currently in use by the
generator.
After selecting the Configure button, the highlighting is cleared because the displayed values now
match the newly configured generator settings.
When therapy is turned on or setting changes are made with the therapy on, the generator needs
approximately 3 minutes to resynchronize with respiration. For this reason, it is recommended to make
changes only after observing therapy performance for 3–5 minutes.