Programmer's Manual

Table Of Contents
Programmer Model 2740 English 57
Delete a session report Highlight report and select the Delete button to display the Delete Report screen
(Figure 4-20).
Select the Delete Selected Report option to remove current report, or
Select the Delete All Reports Before option and specify the date using the
arrow buttons to remove multiple reports for the current patient.
Select Cancel button to return to Reports screen.
Select Delete button to confirm that you wish to permanently remove the
selected report(s).
Export session report to
external USB flash drive
Attach the Inspire USB flash drive to the tablet.
Highlight report and select the Export button to display Export Report screen
(Figure 4-22). If Export button is disabled, go to Programmer Settings screen
to enable USB export.
Select Export Selected Report button to export current report, or
Select Export Combined Visit Report button to export a report that combines
programming sessions from the same visit, or
Note: The option to export combined reports will only be available if multiple
reports are detected within a +/-24-hour period.
Select Export All Reports Since button and specify the date using the arrow
buttons to export multiple reports for the current patient.
Select the information to be included in the export by checking one or more
of the following boxes: patient information, device information, waveform
screenshots, and advanced settings.
Note: Session reports will be exported as PDF (portable document format) files.
The programming log contains a CSV (comma separated value) file of all
generator setting changes for the session and a CSV file of the raw waveform
data.
Select Cancel button to return to Reports screen.
Select Export button to complete export.
Table 4-5. Reports Screen Buttons