Programmer's Manual

Table Of Contents
Programmer Model 2740 English 67
Retain Records
Export a session report if desired.
Initial Activation
The goal during Initial Activation is to program the patient with stimulation and timing settings that will promote
effective and comfortable use of Inspire therapy.
Initial Activation Overview
Start Session
Start Initial Activation Workflow
End Session
Retain Records
Start Session
Turn on the programmer and start a programming session by connecting to the generator.
Start Initial Activation Workflow
From the Home screen select Start for the Initial Activation workflow.
Determine Stimulation Levels
Therapy must be turned off prior to starting the Initial Activation workflow.
1. Use the arrow buttons to select the test stimulation amplitude value.
2. Select the Test Stimulation button to deliver one burst of stimulation at the selected amplitude; observe
the patient’s tongue response.
3. Repeat steps 1 and 2 until the patient can feel stimulation.
4. Select the Save button for the Sensation level to record the amplitude value.
5. Repeat steps 1 and 2 until a functional tongue motion is observed.
6. Select the Save button for the Functional level to record the amplitude value.
7. Select a description of the patient’s tongue motion at the Functional level.
8. Select the Next Step button.