68 English Programmer Model 2740
Figure 5-7. Initial Activation - Determine Stimulation Levels Screen
Observe Waveform
1. It is recommended to start with the functional level. Select the Set Amplitude button.
2. Select the Start Waveform button and observe the waveform for 3 minutes.
Note: This will turn on therapy and deliver stimulation based on the current generator settings.
3. If stimulation is uncomfortable for the patient, decrease the amplitude and select the Set Amplitude
button.
4. If necessary, repeat step 3 until a comfortable amplitude is achieved.
5. Select the Screenshot button to annotate and save the displayed waveform in the session report.
6. Select the Next Step button.
Note: This will stop stimulation and the real-time waveform.