Programmer Model 2740 English 75
4. Select Configure to program the new pulse width / rate and amplitude if the desired settings are new.
5. Select the Next Step button.
Figure 5-16. Advanced Follow Up - Select Pulse Width and Rate Screen
Settings Summary
1. Review the recommended settings.
2. If desired, make additional changes using the arrows and select the Test Stimulation button to evaluate.
3. If there are new settings to program in the generator, select the Configure button.
4. Select the Done button.
End Session
Before ending the Follow Up session:
1. Confirm that therapy is turned off.
2. Review the Settings Summary and confirm desired changes.
3. Confirm that the Follow Up Workflow has been completed.
4. Select the Exit button on the Home screen.