If programmer screens do not display the correct number format, use the Programmer Settings screen to change
the number format.
Therapy
Setting Changes Not in Generator
If setting changes made during a session do not appear on the Home screen, in a session report, or in the
generator:
1. Return to the appropriate programming screen and repeat the desired setting changes.
2. Push the Configure button. If you leave a screen before pushing this button, the changes will not be saved
to the generator.
3. Review settings on the Home screen before ending the session.
4. Always allow all communications to complete before a patient leaves. If a patient is allowed to leave during
a test stimulation or impedance measurement, the generator may not be restored to therapy settings.
Usage
The therapy usage status displays red on the Home screen when therapy has been on for less than 4 hours per
night since the last programming session. Reference the last session date on the Home screen.
Therapy Is On But Stimulation Is Not Active
If you have selected the Therapy On button but therapy appears to be inactive, the generator may be in start delay
or pause time.
Patient Amplitude Change Is Incorrect
If the previous programming session was not ended properly or was conducted with an earlier version of the
programmer software, the patient amplitude change information that displays on the Home screen may be
incorrect.
To ensure that patient amplitude change information is correct follow these guidelines:
Always review therapy settings and properly end the programming session before the patient leaves.
(See "Exiting a Session and Turning Off the Programmer" on page 29 for more information.)
Never let a patient leave after a forced programmer shutdown until you reconnect to the generator and
confirm the generator settings.