Programmer's Manual

Table Of Contents
Programmer Model 2740 English 87
Jerky or Halting Stimulation
Jerky or halting stimulation can occur as a result of connection interrupting stimulation. To reduce the likelihood of
halting stimulation:
1. Turn off therapy before testing stimulation or changing generator settings.
2. Reposition programming head to achieve stronger signal strength.
Measure Impedances
The Measure Impedances screen (Figure 6-4) can be accessed from the System Check workflow. This screen is used
to assess the integrity of leads, lead electrodes, and the lead-generator connection.
Figure 6-4. Measure Impedances Screen
Follow the numbered steps on screen to measure impedances:
1. Review Amplitude.
Use the arrow buttons to increase or decrease amplitude.
The default amplitude setting is 1.5 V. Higher amplitudes increase measurement accuracy, but may be
uncomfortable for the patient.
2. Select the Measure Impedances button.
It is recommended to measure impedances at 1.5, 2.0, 2.5, 3.0, and 3.5 V while the patient remains
comfortable.
3. Analyze impedance measurement results.
A measured impedance < 200 ohms may indicate a short between lead conductors preventing
stimulation from reaching the patient.
Additionally, certain combinations of impedance results may indicate a problem. These will be flagged
as abnormal if they occur.
Impedance values should be considered for information only.