Electromagnetic Emissions
This equipment is intended for use in the electromagnetic environment specified below. The user of this equipment should assure
that is used in such an environment.
Emissions Compliance according to Electromagnetic environment
RF Emissions (CISPR 11) Group 2 The OmniPod and the PDM emit low level electromagnetic energy
(RF) in order to communicate. Although unlikely, nearby electronic
equipment may be affected.
CISPR B
Emissions Classification
Class B The OmniPod is suitable for use in all establishments including
domestic establishments.
Electromagnetic Immunity
The OmniPod is intended for use in the electromagnetic environment specified below. You should observe these requirements in
the use of the OmniPod System.
Immunity against
IEC 60601-1-2
test level
Compliance level
(of this device) Electromagnetic environment
ElectroStatic Discharge,
ESD (IEC 61000-4-2)
contact discharge:
± 6 kV
air discharge: ± 8 kV
± 8 kV
± 15 kV
If floors are covered with synthetic material, try to avoid electro-
static discharges.
Power Frequency mag-
netic fields 50/60 Hz
(IEC 61000-4-8)
3 A/m 400 A/m Suitable for most environments. Magnetic field strengths in
excess of 400 A/m would be unlikely except in close proximity
to industrial magnetic devices.