User Manual

Table Of Contents
Checking Your Blood Glucose
89
7
Use only one edge of the blood glucose test strip per test. Do
not
apply blood to both edges (Figure 7-27). Only use strips
once. Discard used strips.
The test strip pulls the blood into the strip through the edge
(Figure 7-28).
Wait for the reading
1. Do not pull the test strip away from the sample until you hear
1 beep or see the word Checking on the screen (Figure 7-29,
on the next page). This means you applied enough blood
and the meter is reading your glucose.
The lancing device is intended for individual use and
should not be shared. If the device is being operated
by a second person who is providing testing assis-
tance to the user, the meter and lancing device
should be decontaminated prior to use by the sec-
ond person (see Page 138).
Figure 7-27
Edge of
test strip
Figure 7-28
Do not:
Press the strip against the test site
Scrape the blood onto the strip
Apply blood to the flat side of the strip
Apply blood to the strip when it is out of the meter
Put blood or foreign objects into the test strip port