174
18Browsing Your History and Records
18.5. Calculations for History Summaries
is section lists the calculations for the summary data shown on the history
screens.
Glucose summaries
e calculations used for the glucose summaries include CGM values (including
HIGH and LOW values) and exclude manually-entered readings.
Item Calculation
Row 1
Average
CGM
=
Sum of all CGM values
Total number of CGM values
Note: HIGH CGM values are included as 401 and LOW CGM
values as 39.
CGM in
Range
=
Qty of CGM values within CGM/BG Goal Range
x 100
Total number of CGM values
CGM
Above
=
Qty of CGM values above CGM/BG Goal Range upper limit
x 100
Total number of CGM values
CGM
Below
=
Qty of CGM values below CGM/BG Goal Range lower limit
x 100
Total number of CGM values
Insulin delivery summaries
Bolus insulin calculations include Bolus Calculator boluses and manual boluses.
If you cancel an immediate or extended bolus before it completes, only the
amount actually delivered is included in the calculation.
Basal insulin calculations include insulin delivered according to the Basal
Program in progress adjusted for periods when a temp basal was in progress,
insulin was paused, or there was no active Pod.
When your Omnipod 5 app has not received conrmation from the Pod about
actual insulin delivery, the insulin delivery calculations are estimates based on the
scheduled insulin delivery.
Settings
History
Updates
Effective Date: 22DEC2020, PCO-000267