Users Manual

132
13About the DexcomG6
13.6. Communication Messages
e DASHBOARD displays the following communication messages:
CONNECTING TRANSMITTER:
Occurs aer you have entered in a
transmitter serial number and the
Pod is attempting to connect with
the transmitter.
WAITING FOR DEXCOM SETUP:
When the transmitter is connected
but CGM values are unavailable
because the
Dexcom G6 is in sensor
warm-up or requires calibration. See the
Dexcom G6 app for details. No
action is required within the Omnipod 5 app.
SEARCHING FOR POD: When Pod communication was not established
within the most recent 5-minute update interval. Tap MORE
INFORMATION for potential causes and recommended actions.
SEARCHING FOR CGM: When the CGM is active and connected to the
Omnipod 5 Pod but the most recent CGM value was not acquired within the
5-minute window. ere may be no valid CGM value available due to a Pod/
CGM communication issue or a temporary CGM sensor issue (recoverable
without any user action). Tap MORE INFORMATION for recommended
action. Review Pod and CGM placement. Pod and CGM should be within
line of sight.
DEXCOM ISSUE DETECTED: When
CGM values are not available due
to a sensor error (including sensor
expiration). See the
Dexcom G6 app
for details. No action is required
within the Omnipod 5 application.
TRANSMITTER ERROR: When the
transmitter that was connected with the Omnipod 5 System has expired or
experienced a non-recoverable error. Tap NEED HELP for potential causes
and recommended actions. To set up a new transmitter, see "Chapter 12:
Connecting CGM to the Pod" on page 123.
TRANSMITTER NOT FOUND: When the Pod has tried to connect with a
transmitter but aer 20 minutes was unable to do so. Tap NEED HELP for
potential causes and recommended action.
Note: For all CGM related issues, refer to your compatible CGM User Guide.
Using a
CGM with
OmniPod 5
Effective Date: 22DEC2020, PCO-000267