10 Alarms, Notications, Communication Errors
If the PDM has sent the Pod the instruction and hasn't received conrmation that
it was carried out, the PDM oers these options:
CHECK STATUS: Move to a new location, then select this option to recheck for
conrmation that the instruction was carried out.
DEACTIVATE POD: Select this option and follow the PDM's instructions for
replacing your Pod.
If the PDM has not sent the Pod the instruction, the PDM tells you to move to a
new location and tap TRY AGAIN to reattempt communication. Aer you tap TRY
AGAIN, if the next communication attempt fails, the PDM oers these options:
CANCEL: Select this option to cancel sending the instruction. In this case, the
Pod continues with its prior insulin delivery mode. You can try to send the
instruction later.
TRY AGAIN: Move to a new location, then select this option to tell the PDM to
reattempt to send the instruction to the Pod.
DEACTIVATE POD: Select this option and follow the PDM's instructions for
replacing your Pod.
Note: At any point, if the PDM receives conrmation from the Pod that the
instruction was successfully received, a green banner appears informing you that
the instruction has been carried out.
Error when activating a Pod
If a communication error occurs during Pod activation, the PDM oers these
options:
DISCARD POD: Select this option to stop attempting to use this Pod.
TRY AGAIN: Select this option to attempt to reestablish communication.
Error when deactivating a Pod
If a communication error occurs during Pod deactivation, the PDM oers these
options:
DISCARD POD: Select this option to tell your PDM to unpair from that Pod.
e PDM instructs you to remove your Pod and tap CONTINUE.
TRY AGAIN: Select this option to attempt to reestablish communication.
Note: Aer selecting the discard option, you can prevent future alarms from the
discarded Pod by following the instructions in "Silencing an Alarm" on page 135.
Note: If there is an unconrmed bolus when you discard a Pod, the PDM does
not know how much of the bolus was delivered. erefore, the PDM temporarily
disables the Bolus Calculator. If you tap the Bolus button while the Bolus
Calculator is disabled, the PDM displays a message that says "Bolus Calculator