Storage atmospheric pressure: 700 hPA to 1060 hPA
Non-pyrogenic: Fluid pathway only
Type BF applied part: Protection from electrical shock
Maximum infusion pressure: 35 psi
Maximum volume infused under single fault conditions: 0.0 U
Flow Capability:
Basal: Programmable by the user in 0.05U increments up to 30.0 U per hour
Bolus Rate: 1.5 units per minute. Dose range from 0.05 to 30.0 units
Delivery accuracy (tested per IEC 60601-2-24):
Basal: ± 5% at rates ≥ 0.05 U/hr
Bolus: ± 5% for amounts ≥ 1.0 unit
± 0.05 units for amounts < 1.0 unit
Note: e user should consider bolus dose accuracy when setting a bolus
dose. When using the lowest bolus dose allowable (0.05 units), the actual bolus
delivered may be as low as 0.00 units or as high as 0.10 units.
Accuracy test results: e following graph shows the ow accuracy of the Pod
against given time periods. e measurements were made using a Pod with a
basal rate of 0.5 μl/h (which delivers 0.05 U/h of U-100 insulin) at a high operating
temperature. e overall mean percentage ow error was 1.40%.
Observation Interval (min)
Error (%)
EpMax
EpMin