PDM Specications
Size: 2.48” wide x 5.1” long x 0.39” high (6.3 cm x 13.0 cm x 1.0 cm)
Weight: 6.17 oz (175 grams)
Screen active area: 4.0" diagonal ± 5% (10.2 cm ± 5%)
Operating temperature range: 41°F to 104°F (5°C to 40°C)
Storage temperature range: 32°F to 86°F (0°C to 30°C)
Operating relative humidity range: 20% to 90%, non-condensing
Storage relative humidity range: 20% to 90%, non-condensing
Operating atmospheric pressure: 700 hPA to 1060 hPA
Storage atmospheric pressure: 700 hPA to 1060 hPA
Communication distance: e PDM and Pod should be
At startup: Adjacent and touching, either in or out of tray, to ensure proper
communication during priming.
During normal operation: Within 5 feet (1.5 m) of each other. Depending on
the location, the communication distance may handle separations up to 50
feet (15 meters) away.
Waterproof rating: IP22 (avoid liquid)
Alarm type: Audible. Output: ≥ 45 db(A) at 1 meter
Notication type: Audible and vibratory
Battery: Rechargeable Li-ion battery, 3.7V, 1300 mAh
Only use the NUU mobile Li-Ion battery Model NUBA1 (Insulet PN 18363)
with the PDM.
Battery life: Full charge covers approximately 2 days of typical use aer 2 years of
typical use
Battery charger operating line voltage: 100 to 240 VAC, 50/60 Hz
Only use the NUU mobile Switching Adapter Model HJ-0501000E1-US
(Insulet PN 18401) with the PDM.