Operator's Manual

Table Of Contents
Live Case Export
BIS™ Advance Monitor
Download of Live Case data cannot commence unless a USB flash drive is
connected to the monitor or the docking station.
2. Live Case data can be exported during monitoring only, only when a USB
flash drive is connected to the monitor (either directly to the monitor or via
the docking station). Commence monitoring.
3. Click Main Menu>DOWNLOAD. Select Live Case. The left button in the
middle of the screen will display a grey circle. Click this button , so that it
will display a red circle and the word REC in red will appear next to Live
Case at the left side of the window.
4. When Live Case recording is on, and a flash drive is connected to receive
the Live Case recording, the Live Case icon will appear at the top right side
of the home screen (reference 3.10.9 Home Screen Indicators and Messages
on page 114).
5. Live Case data will begin recording onto the USB flash drive. The Live Case
icon at the top right side of the home screen (reference 3.10.9 Home Screen
Indicators and Messages on page 114) will indicate that Live Case recording
is on.
6. Live Case data download will continue until one of the following occurs:
The monitor is shut down
The USB flash drive is removed
Live Case data download is stopped by turning Live
Recording off by clicking the right square in the Live Case
DOWNLOAD window.
The case ends
The operator removes and reinserts or switches the sensor or
the BISx module (which will start a new case).
7. If the operator removes and reinserts or switches the sensor or the BISx
module (which will start a new case), Live Recording must be set to ON
again before download can commence.
8. The Live Recording default setting is OFF. Live Recording should be
set to ON as described above before each new case.