User's Manual

EMC Information
Chapter 8
Page 130
8.5 Immunity and Compliance Level
Immunity test
IEC 60601 Test
Level
Actual Immunity
Level
Compliance
Level
Conducted RF
IEC 61000-4-6
3Vrms
150kHz to 80MHz
3Vrms 3Vrms
Radiated RF
IEC 61000-4-3
3Vrms
80MHz to 2.5GHz
3V/m 3V/m