User's Manual

Safety Information Chapter 1
1- 14
1.4.1 Environmental Conditions for Storage
Temperature : -10 ~ 70 (Operating 0 ~ 40 )
Relative humidity : 10 % ~ 80 % (Operating 45 % ~ 75 %)
Atmospheric pressure : 700hPa to 1060hPa
WARNING
DO NOT operate the equipment in the vicinity of
generators, power stations, X-ray devices, and
broadcasting stations where high levels of electro-magnetic
waves are generated. The electro-magnetic waves can
cause equipment malfunctions.
CAUTION
If the equipment has been brought in from a cold
environment (stock room, airfreight) into a warm room,
initial activation should take place after a few hours, to
allow for temperature adjustment and balance and
evaporation of condensed humidity.
WARNING
DO NOT operate the equipment in the vicinity of heat
sources, strong electric or magnetic fields (close to a
transformer), or near instruments generating
high-frequency signals.
WARNING
DO NOT use MiroCam® alongside or together with medical
devices or procedures involving electrical currents.
DO NOT use MiroCam® with h.f. surgical equipment. It
may result in burns at the site of the electrodes and
possible damage to the MiroCam® Capsule and the
MiroCam® Receiver.
DO NOT use the unit in close radius (within 1 m) of short
wave or microwave therapy equipment. It may produce
instability in the captured image.
WARNING
This device is a Class B device according to EN60601-1-2
standards. This equipment can cause radio interference in
residential areas. In this case, the owner (or operator) can
be held responsible to take appropriate measures or take
proper measures for compensation.