User's Manual
Performing Capsule Endoscopy Chapter 3
3- 4
CAUTION
If SIG indictor of the MiroCam® Receiver does not turn
green when the patient holds the MiroCam® Capsule by
the gold bands, immediately stop the MiroCam® Capsule
endoscopy procedure.
NOTE
If images of the small bowel are not captured during the
MiroCam® Capsule endoscopy procedure, ensure the
MiroCam® Capsule has been excreted by the patient and
perform procedure again.
3.3 Examination Preparation
The following steps (3.3.1 to 3.3.4) explain the process to initialize the
MiroCam® Receiver and register patient data for the MiroCam® Capsule
endoscopy procedure. For more information, see section 4.2.
3.3.1 Start System
Turn on the power of the workstation.
Click the
to start the 0LUR9LHZOperator.
Type the ID and the password in each field, and then click the
.
WARNING
Before starting the system, check the power rating for the
workstation.