User's Manual Part 1

Introduction
da Vinci
®
Si™
1-6
DRAFT/PRE-RELEASE/CONFIDENTIAL
10/9/14
The integrated power strip in the Vision Cart also has the label below:
Figure 1.4 Vision Cart integrated power strip caution label
A label wrapped around the Vision Cart power cord also reinforces this point:
Figure 1.5 Vision Cart power cord caution label
The Surgeon Console, Patient Cart, and Vision Cart automatically adapt to 100VAC, 120VAC, or
230VAC. (This is called “Auto Sense” in the table below.) Please refer to the electrical rating
label located on the bottom rear panel of the da Vinci Si System components.
For Intuitive Surgical Use Only.
150 VA maximum power available from unused
receptacles on the power strip.
100/120/230V~
50/60 Hz
150 VA Continuous”
Caution: Vision Cart requires dedicated
circuit. Do not connect auxiliary equipment to
the circuit powering the Vision Cart.
System Component Voltage Rating
Surgeon Console
100/120/230V~
50/60Hz
Auto Sense
1000VA Continuous
Patient Cart
100/120/230V~
50/60Hz
Auto Sense
1000VA Continuous
Vision Cart
includes Core and
Camera Control Unit
100/120/230V~
50/60Hz
Auto Sense
1500VA Continuous
Core
a,b
100/120/230V~ or 100-230V~
50/60Hz
Auto Sense
650VA Continuous
Camera Control Unit
(CCU)
a
100/120/230V~
50/60Hz
Auto Sense
100VA Continuous
a. Core and Camera Control Unit (CCU) power requirements are included in the Vision Cart.
These are provided separately for reference only.
b. If the Core is powered separately from the Vision Cart, it must be powered from a cen-
ter-tapped, 240V, single phase circuit. This applies to the US only.