User's Manual Part 1

da Vinci
®
Si™
System Connections
4-5
DRAFT/PRE-RELEASE/CONFIDENTIAL
10/9/14
How to Connect System Cables
1. Remove the cables' protective caps, inspect the cable connectors and system receptacle
for debris or bent pins. The system receptacles have protective covers you must open.
2. Connect the system cables between the components as shown in Figure 4.3. To connect
the cables, line up the red dot on the cable c
onnector with the red dot on the matching
receptacle, flip up the receptacle cover and insert the connector. You should hear an
audible click when the cable engages properly. Gently pull on the connector to verify the
cable is fully seated.
Figure 4.4 Connecting system cable to Surgeon Console
Note: The protective metal caps attached to each system cable should be installed on
cable
ends a
t all times when not connected to the system.
Figure 4.5 System connections
Red alignment mark
1. Observe red mark
and align
2. Flip up receptacle
cover & insert
3. Tug gently to assure
connection
Patient Cart connection
Vision Cart Core connections
Surgeon Console connection