User's Manual Part 1

da Vinci
®
Si™
System Connections
4-13
DRAFT/PRE-RELEASE/CONFIDENTIAL
10/9/14
If an upgrade has been installed that makes available the optional TilePro Input bays on
the Core (see Figure 4.15), you can use either set of TilePr
o inputs, but not more than two.
The system automatically checks for TilePr
o connections in order: first on the Core (if
present), then on the Surgeon Console. The Core TilePro inputs take precedence if they
are in use.
Core Video Outputs
The Core supports up to three video output bays (Figure 4.15). Video output from all three
includes the icons and text messages displayed in the Surgeon Console. Each video output
bay supports a single DVI, HD-SDI, S-Video, or Composite signal.
Note: Video Out bays 1 and 2 are not available by default; they are made available in
conjunction with optional upgrades, for example, with purchase of a second console to
support dual console surgery. Only the Video Out Aux bay is available by default.
Core Audio Inputs and Outputs
Three connectors are provided on the audio bay (Figure 4.15).
2.5 mm input/output headset jack
•RCA input
•RCA output
Camera Control Unit Video Outputs
Two video outputs are available on the back of the Camera Control Unit. The left-eye (L) and
right-eye (R) video outputs provide the surgical image without icons or text messages. This is
component video, made up of Y (green port), P
R
(red port) and P
B
(blue port).
Figure 4.17 Video connections (back of CCU)
Troubleshooting: Audio not Functioning
1. Power off the system.
2. Check audio cables for debris or bent pins.
3. Reconnect all audio cables.
4. Power on the system.
a. Check audio on the Vision Cart and Surgeon Console. Adjust volume if needed.
5. If the problem persists, contact customer ser
vice f
or assistance: U.S. 800-876-1310,
International +800 0821 2010 or +41 21 821 2020.