User's Manual Part 1

Startup
da Vinci
®
Si™
5-3
DRAFT/PRE-RELEASE/CONFIDENTIAL
10/9/14
2. Confirm that the mains circuit breaker switch on each of the Surgeon Console, Patient
Cart and Vision Cart is correctly set to the on position (indicated by “I” near each switch).
These switches are found on the rear of each component near the power cord inlet.
Figure 5.2 Circuit breaker switches
on (“I”) for main components
On the Patient Cart, make sure that the EPO button is not depr
essed, since it requires
a second press after use to reset it once it is used. See “Hard-Cycle” Power below.
3. On the Vision Cart, confirm that each of its component power switches—for the Core,
Camera Contr
ol Unit and Illuminator—is correctly set to the on position (indicated by “I”
on the switch). These components are not intended to be turned off or disconnected at
any time unless you are directed to do so by Intuitive Surgical Product Support. Also
confirm that each Vision Cart component power cord is plugged into the integrated
Vision Cart power strip.
Figure 5.3 Vision Cart comp
onent switc
hes on (“I”)
Check Connection of System Cables
Check that the blue system cables are properly connected between the Core and the Surgeon
Console and Patient Cart. See section 4.2 System Cable Connections on page 4-3.
“Hard-Cycle” Power
If the entire system or one of the Patient Cart, Vision Cart or Surgeon Console fails to power on
normally by pressing one of the Power buttons, first check the AC power connections as
described above. If all AC power connections are complete, follow these steps to “hard-cycle
power on the whole system:
Patient Cart
Vision Cart (at base)
Surgeon Console
EPO button not depressed
Illuminator
Camera Control Unit
Core