User's Manual

Da Vinci IS5000 System User Manual 555500-01 Rev. A
The integrated power strip in the Tower also has the label below:
Figure E.2 Tower integrated power strip caution label
For Intuitive Use Only.
150 VA maximum power available from unused receptacles
on the power strip.
100/120/230 V~
50/60 Hz
150 VA Continuous
A label wrapped around the Tower power cord also reinforces this point:
Figure E.3 Tower power cord caution label
Caution: Vision Cart requires dedicated
circuit. Do not connect auxiliary
equipment to the circuit powering the
Vision Cart.
The tables below provides dedicated circuit requirements for each component.
Table E.1 Dedicated circuits required for 115 VAC, 60 cycle at 15 and 20 amperes
System Component Circuit Voltage
VAC
Ampere
Service
Console 1 1 115 15
Optional
Console 2 2 115 15
Robot 3 115 15
Tower 4 115 20
a
This circuit can be shared with other auxiliary devices
such as an insufflator, third-party electrosurgical unit,
video recorder, etc.
5 115 As needed
Notes
a. Circuit breaker for the Tower needs to be suitable for motor-transformer loads. (For example,
tripping characteristic “K”.)
E-2
System Specifications | Power Requirements – Main Components