E.8 Environmental Specifications
Table E.10 Environmental conditions: operating
Environmental Conditions: operating
Temperature 10 to 30 °C (50 to 86 °F)
Humidity 10 to 85% non-condensing
Atmospheric Pressure The system shall function properly in atmospheric pressures ranging
from 523 mm HG (10,000 ft.) to 774 mm HG (-500 ft.). For every 1000
feet above sea level, the 30 ºC operational temperature limit specified
above will be reduced by 1 ºC. (For example, the maximum operating
temperature at 5000 feet will be 25 ºC, and the maximum operating
temperature at 10,000 feet will be 20 ºC.)
Table E.11 Environmental conditions: storage and transport
Environmental Conditions: Storage and Transport
Temperature -20 to 60 ºC (-4 to 140 °F)
Humidity 5 to 90% non-condensing
Note: Storage and transport of the system are not affected by atmospheric pressure.
E.9 Crate Dimensions
Table E.12 Crate dimensions
L x W x H Weight
Console ?x?x?"
(? x ? x ?m)
? lbs (? kg)
Robot 70 x 50 x 87.5"
(1.78 x 1.27 x 2.22 m)
2292 lbs (1039.6 kg)
Tower ?x?x?"
(? x ? x ?m)
? lbs (? kg)
E.10 Electromagnetic Compatibility
The system has been tested and found to be in compliance with IEC 60601-1-2:2007 or IEC
60601-1-2:2014, depending on the date of manufacture. For more information on IEC
compliance, contact Intuitive Customer Service.
It is intended to be used in an operating room environment and operation of the system is
unaffected when used in the electromagnetic environment as specified in Tables 2 4 in the EMC
System Specifications | Electromagnetic Compatibility
E-7