5 MAINTENANCE AND TROUBLESHOOTING
Invacare
®
APM
Model CG9900 Mattress 26 Part No 1171739
10. Perform the following steps to clean the mattress components (air cushions, mattress base, etc.):
A. Using a clean sponge or paper towel, wipe down the mattress components with a dilute detergent solution of quaternary cleaner disinfectant
or other germicidal detergent solution.
B. Wipe the mattress components with a clean dry cloth.
Storing the System
1. Ensure that the control unit is off and disconnect the power cord from the wall outlet.
2. Clean the system. Refer to Cleaning the System on page 24.
3. Disconnect the air hose connector(s) from the control unit and allow air to vent from the mattress.
4. Gently roll up the mattress with minimal handling and agitation.
5. Ensure the cover surface is inside the roll.
6. Store the unit, keeping the mattress with the control unit.
Replacing Bladders
To replace an individual bladder:
1. Unsnap the buttons for the bladder on both sides.
2. Cut a small slit on the red female connector.
3. Pull apart the cut female end of the connector and bladder from the male end of the manifold.
4. Take the new bladder and dampen the inside of the female end of the connector.
5. Push the new female end of the connector and bladder onto the male end of the manifold securely.