Product Manual
2 OVERVIEW
Part No 1171739 7 Invacare
®
APM
™
Model CG9900 Mattress
2.2 Typical Product Parameters
Electrical Parameters
Performance Parameters
CG9900
INPUT VOLTAGE AC: 90 V
INPUT FREQUENCY: 60 Hz
CURRENT: 1 A
MAXIMUM POWER CONSUMPTION: 5 ± 3 W
CIRCUIT PROTECTION: Dual fused, 250 V, 1 A fast blow fuses
MODE OF OPERATION: Continuous
CG9900 CG9900
WEIGHT CAPACITY:
PRESSURE ZONE:
MAXIMUM FLOW:
MAXIMUM FLOW PRESSURE:
MAXIMUM FLOW TIMER:
SUPPORT SURFACE INFLATION
TIME:
350 lbs
2
12 ± 4 LPM
35 ± 5 mmHg
45 minutes
20 - 45 minutes
PATIENT COMFORT
CONTROL
PRESSURES
SOFT PRESSURE:
FIRM PRESSURE:
CYCLE TIME:
8 ± 4 mmHg
32 ± 4 mmHg
10 minutes
PATIENT CONTACT:
Control unit and mattress
have Latex-Free components