Operator’s Manual Platinum™ Series XL, 5, 10 Oxygen Concentrators Standard, with SensO2®and HomeFill® Invacare 5 with SensO2 Model IRC5LXO2 Invacare 5 Model IRC5LX Invacare 10 with SensO2 Model IRC10LXO2 Invacare 10 Model IRC10LX Dealer: This manual MUST be given to the end user. User: Before using this product, read this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.invacare.
WARNING WARNING DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner’s manuals, service manuals or instruction sheets supplied with this product or optional equipment.
TABLE OF CONTENTS TABLE OF CONTENTS ACCESSORIES .................................................. 2 REGISTER YOUR PRODUCT .............................. 4 SPECIAL NOTES ............................................... 5 LABEL LOCATION ............................................ 7 SECTION 1—GENERAL GUIDELINES .................. 8 To Reduce the Risk of Burns, Electrocution, Fire or Injury to Persons ................................................................................. 8 Operating Information ..
TABLE OF CONTENTS TABLE OF CONTENTS SECTION 6—MAINTENANCE .......................... 23 Routine Maintenance........................................................................ 23 Cleaning the Cabinet Filter......................................................... 23 Cleaning the Cabinet ................................................................... 24 Cleaning the Humidifier .............................................................. 24 SECTION 7—TROUBLESHOOTING GUIDE ......... 26 NOTES ......
SPECIAL NOTES SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words. SIGNAL WORD MEANING DANGER Danger indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. WARNING Warning indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
SPECIAL NOTES CAUTION “Caution: Federal law restricts this device to sale or rental by or on order of a physician, or any other practitioner licensed by the law of the State in which he/she practices to use or order the use of this device.” Invacare recommends an alternate source of supplemental oxygen in the event of a power outage, alarm condition or mechanical failure. Consult your physician or equipment provider for the type of reserve system required.
LABEL LOCATION LABEL LOCATION IRC5LXO2 and IRC10LXO2 IRC5LX and IRC10LX HomeFill II Compatib Compatible le HomeFill II Compatible with Operating Instructions Alarms SensO2™ Operating Instructions (See Operator's Manual) (See Operator's Manual) I/O Alarms (See Operator's Manual) (See Operator's Manual) WARNING - RISK OF FIRE WARNING - RISK OF FIRE NO SMOKING - Keep ALL sources of ignition out of the room in which this product is located and away from areas where oxygen is being delivered.
SECTION 1—GENERAL GUIDELINES SECTION 1—GENERAL GUIDELINES In order to ensure the safe installation, assembly and operation of the Platinum concentrator these instructions MUST be followed. WARNING This section contains important information for the safe operation and use of this product. To Reduce the Risk of Burns, Electrocution, Fire or Injury to Persons DANGER Risk of electric shock. DO NOT disassemble. Refer servicing to qualified service personnel. No user serviceable parts.
SECTION 1—GENERAL GUIDELINES Operating Information For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time. Shorter periods of operation may reduce maximum product life. Keep the oxygen tubing, cord, and unit out from under such items as blankets, bed coverings, chair cushions, clothing and away from heated or hot surfaces, including space heaters, stoves and similar electrical appliances.
SECTION 1—GENERAL GUIDELINES Other devices may experience interference from even the low levels of electromagnetic emissions permitted by the above standards. To determine if the emissions from the Platinum is causing the interference, turn the Platinum Off. If the interference with the other device(s) stops, then the Platinum is causing the interference.
SECTION 2—FEATURES SECTION 2—FEATURES Oxygen Purity Indicator Lights (SensO2 model only)/ Fault and Power Indicator Lights (LX model only) Oxygen Outlet Flowmeter Power Switch Circuit Breaker Elapsed Time Meter Cabinet Filter (one of two shown) ACCESSORIES (NOT SHOWN): Remote Humidifier/Flowmeter Accessory IRCRF5 (Platinum 5 and XL Only) REAR VIEW PreciseRX™ Pediatric Humidifier/Flowmeter Accessory - IRCPF16 (Platinum 5 and XL Only) HomeFill home oxygen compressor - IOH200 Power Cord Part No 1118353
SECTION 3—HANDLING SECTION 3—HANDLING The concentrator should ALWAYS be kept in the upright position to prevent cabinet damage while being transported. If the concentrator is to be reshipped by common carrier, additional cartons are available from Invacare. Unpacking NOTE: For this procedure, refer to FIGURE 3.1. 1. Check for any obvious damage to the carton or its contents. If damage is evident, notify the carrier, or your local dealer. 2. Remove all loose packing from the carton. 3.
SECTION 4—TYPICAL PRODUCT PARAMETERS SECTION 4—TYPICAL PRODUCT PARAMETERS Electrical Requirements: 120 VAC ± 10% (132 VAC/108 VAC), 60 Hz Rated Current Input: Platinum 5 and XL: 4.3 A Platinum 10: 4.7 A Sound Level: Platinum 5 and XL: 50 dB Average Platinum 10: 58 dB Average Altitude: Platinum 5 and XL - Up to 6,000 ft (1828 meters) above sea level without degradation of concentration levels.
SECTION 4—TYPICAL PRODUCT PARAMETERS Flow Range: Platinum 5 and XL - 0.5 to 5 L/min For flowrates less than 1 L/min, we recommend the use of the Invacare Pediatric Flowmeter IRCPF16 Accessory. Platinum 10 - 2 L/min to 10 L/min Low Flow Alarm: Platinum 5 - 0 L/min to 0.5 L/min, Platinum 10 - 0 L/min to 1 L/min Rapid Audible Alarm Beeping (No Accessories Connected) Average Power Consumption: Platinum 5 and XL: 400W Platinum 10: 585W Pressure Relief Mechanism Operational at: 35 psi ± 3.
SECTION 4—TYPICAL PRODUCT PARAMETERS Operating Ambient Temperature: 50°F - 95°F (10°C - 35°C) at 20-60% relative humidity Exhaust: Platinum 5 and XL: Less than Ambient + 35°F (+19°C). Platinum 10: Less than Ambient + 45°F (+25°C) Oxygen Output Temperature: Platinum 5 and XL: Less than Ambient + 4°F (+2°C) Platinum 10: Less than Ambient + 8°F (+4.
SECTION 5—OPERATING INSTRUCTIONS SECTION 5—OPERATING INSTRUCTIONS Introduction Your oxygen concentrator is intended for individual use indoors. It is an electronically operated device that separates oxygen from room air. It provides high concentration of oxygen directly to you through a nasal cannula. Clinical studies have documented that oxygen concentrators are therapeutically equivalent to other types of oxygen delivery systems. Your provider will show you how to use your oxygen concentrator.
SECTION 5—OPERATING INSTRUCTIONS Set-Up Plug in Power Cord Plug in power cord to an electrical outlet. POLARIZED PLUG WARNING As a safety feature, this appliance may have a polarized plug (one blade is wider than the other). This plug will fit in a polarized outlet only one way. If the plug does not fit fully in the outlet, reverse the plug. If it still does not fit, contact a qualified electrician. DO NOT attempt to defeat this safety feature.
SECTION 5—OPERATING INSTRUCTIONS Humidifier Bottle 3. 4. 5. 6. FIGURE 5.2 Humidifier Compartment Remove the filter access door located on the side of the concentrator. Insert a flathead screwdriver in the plate groove on the top edge of the filter access door and gently pry the filter access door open. Refer to Detail “A” in FIGURE 5.3. The humidifier adapter is next to the inlet filter. Pull up and remove the humidifier bottle adapter. Refer to Detail “B” in FIGURE 5.3.
SECTION 5—OPERATING INSTRUCTIONS 7. Attach oxygen tubing from the humidifier bottle to oxygen outlet connector on the oxygen concentrator (FIGURE 5.4). 8. Attach the cannula/patient supply tubing to the humidifier bottle outlet (FIGURE 5.4). 9. After assembly, ensure that oxygen is flowing through the cannula. Oxygen Outlet Connector Oxygen Tubing Humidifier Outlet FIGURE 5.4 Attaching the Oxygen Tubing Power Switch NOTE: For this procedure, refer to FIGURE 5.5 on page 19. 1.
SECTION 5—OPERATING INSTRUCTIONS Flowrate NOTE: For this procedure, refer to FIGURE 5.6 on page 20. 1. Turn the flowrate knob to the setting prescribed by your physician or therapist. CAUTION Never set the flow greater than 5 L/min on the IRC5. Never set the flow greater than 10L/min on the IRC10. WARNING DO NOT change the L/min setting on the flowmeter unless a change has been prescribed by your physician or therapist.
SECTION 5—OPERATING INSTRUCTIONS Front Panel Indicators - Units with SensO2 NOTE: For this procedure, refer to FIGURE 5.7. The SensO2 feature monitors the purity of oxygen generated by the oxygen concentrator. If purity falls below factory preset standards, indicator lights on the control panel will illuminate. OXYGEN PURITY Auto Shut - Down RED Use Backup Call Supplier Below Normal Normal FIGURE 5.
SECTION 5—OPERATING INSTRUCTIONS Front Panel Indicators - Units without SensO2 NOTE: For this procedure, refer to FIGURE 5.8. Auto Shut - Down Use Backup RED Call Supplier Normal FIGURE 5.8 I/O GREEN Front Panel Indicators - Units without SensO2 Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial start warm‐up time (approximately 30 min). O2 purity will build to a maximum during this period.
SECTION 6—MAINTENANCE SECTION 6—MAINTENANCE WARNING The Invacare concentrators are specifically designed to minimize routine preventive maintenance. Only qualified personnel should perform preventive maintenance on the concentrator. Power should be disconnected before beginning preventive maintenance on the Invacare concentrator. NOTE: At a minimum, preventive maintenance MUST be performed according to the maintenance record guidelines.
SECTION 6—MAINTENANCE 2. Clean the cabinet filters with a vacuum cleaner or wash in warm soapy water and rinse thoroughly. 3. Dry the filters thoroughly before reinstallation. Cabinet Filter (located on both sides of concentrator) FIGURE 6.1 Cleaning the Cabinet Filter Cleaning the Cabinet WARNING UNPLUG the concentrator when cleaning. To avoid electrical shock, DO NOT remove cabinet. Clean the cabinet with a mild household cleaner and non‐abrasive cloth or sponge.
Part No 1118353 25 *NOTE: Refer to Preventive Maintenance section of Service Manual. NOTE: 2,160 hours are equivalent to usage 24 hours per day, for 90 days. 4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 6 months. 26,280 hours are equivalent to usage 24 hours per day, 7 days per week, for 3 years.
SECTION 7—TROUBLESHOOTING GUIDE SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: Alarm: SHORT BEEPS, LONG PAUSE PROBABLE CAUSE: Main Power Loss: 1. Power cord NOT plugged in. 2. No power at outlet. SOLUTION: 1. Insert plug into outlet. 2. Inspect house circuit breakConcentrator ers or fuses. If problem recurs, NOT operating, use a different outlet. power switch ON. 3. Tripped circuit breaker. 3. Call Service Provider. Beep......Beep...... Alarm: CONTINUOUS System Failure: 1a. Remove and clean cabinet 1.
SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: YELLOW or RED 1. Low oxygen purity.* LIGHT Illuminated. 2. Kinked or blocked tubing, cannula or humidifier.* Alarm: CONTINUOUS On red light only. * Only applies to 5LXO2 and 10LXO2 SensO2 models. GREEN LIGHT with YELLOW Light Flashing. Part No 1118353 1. Clean or Replace filters. 2. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power OFF for 60 seconds and then turn power back ON. 3.
NOTES NOTES Platinum™ Series 28 Part No 1118353
NOTES NOTES Part No 1118353 29 Platinum™ Series
LIMITED WARRANTY LIMITED WARRANTY NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. This warranty is extended only to the original purchaser who purchases this product when new and unused from Invacare Corporation or a dealer. This warranty is not extended to any other person or entity and it is not transferable or assignable to any subsequent purchaser or owner.
LIMITED WARRANTY For warranty service, please contact Invacare Corporation's Service Department at the toll free number on the back cover. Upon receiving notice of an alleged defect in a product, Invacare Corporation will issue a serialized return authorization.
This product may be manufactured under one or more of the following patents: US 5,906,672 and US 5,917,135. Other patents pending. Invacare Corporation www.invacare.com USA One Invacare Way Elyria, Ohio USA 44036-2125 800-333-6900 Technical Services 800-832-4707 Trademarks are identified by ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.