User Manual HomeFill® Compressor User: Before using the product, read this manual and save for future reference. Dealer: This manual must be given to the end user. For more information regarding Invacare products, parts, and services, please visit www.invacare.
WARNING WARNING DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner’s manuals, service manuals or instruction sheets supplied with this product or optional equipment.
TABLE OF CONTENTS SPECIAL NOTES ............................................... 5 TYPICAL PRODUCT PARAMETERS ................... 7 LABEL LOCATION ............................................ 9 DEALER INFORMATION ................................... 9 HOMEFILL INFORMATION .............................. 10 SECTION 1—IMPORTANT SAFETY INSTRUCTIONS ............................................. 11 Operating Information ..................................................................... 11 Maintenance............
TABLE OF CONTENTS Connecting/Disconnecting the Cylinder to/from the Compressor .............................................................................................. 23 Connecting the Cylinder to the Compressor........................ 23 Disconnecting the Cylinder from the Compressor.............. 26 Turning the Compressor On ......................................................... 28 SECTION 6—INDICATOR LIGHTS .................... 30 SECTION 7—TROUBLESHOOTING ...................
SPECIAL NOTES SPECIAL NOTES Check all parts for shipping damage. In case of damage, DO NOT use. Contact Invacare for further instruction. Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words. SIGNAL WORD MEANING DANGER Danger indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
SPECIAL NOTES DANGER DO NOT SMOKE while using this device. Keep all matches, lit cigarettes or other sources of ignition out of the room in which this product is located and away from where oxygen is being delivered. NO SMOKING signs should be prominently displayed. Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air. Failure to observe this warning can result in severe fire, property damage and cause physical injury or death.
TYPICAL PRODUCT PARAMETERS TYPICAL PRODUCT PARAMETERS Standard Listing: Double Insulated Product ETL Certified as Complying with UL60601-1, First Edition ETL Certified Complying with CAN/CSA C22.2 No.
TYPICAL PRODUCT PARAMETERS Input Pressure Required: 14 - 21 psi Input Flow Required: 2 L/min Input O2% Required: >90% O2 Width: 20 ¼ inches Height: 15 inches Depth: 16 inches Weight: 33 lbs Shipping Weight: 39 lbs No AP/APG: Not suitable for use in the presence of a flammable anaesthetic mixture. Electrical: No extension cords. Placement: No closer than 7.5 cm (3 in) from any wall, furniture, draperies, or furniture to assure sufficient air flow.
LABEL LOCATION LABEL LOCATION WARNING Keep all connections free of oil and grease to avoid violent ignition. See Owner’s Manual before cleaning. P/N 1075220 Rev C - 10/04 WARNING DO NOT REMOVE UNIT COVER This unit contains moving and high pressure components. Manufacturer Service Only. P/N 1075221 Rev B- 10/04 WARNING Keep all connections free of oil and grease. See Owner’s Manual before cleaning. Never oil or lubricate coupling.
HOMEFILL INFORMATION HOMEFILL INFORMATION HomeFill is an accessory device to the oxygen concentrator that enables the patient to refill custom portable oxygen cylinders for personal use. NOTICE The HomeFill compressor is for use with specially designed Invacare Platinum™5, 10, XL and Perfecto2™ HomeFill compatible oxygen concentrators. The oxygen concentrator can be set to the following maximum flowrates when operating with HomeFill compressor: Platinum 5, XL and Perfecto2: 3 L/min or less.
SECTION 1—IMPORTANT SAFETY INSTRUCTIONS SECTION 1—IMPORTANT SAFETY INSTRUCTIONS WARNING This section contains important information for the safe operation and use of this product. DANGER Risk of electric shock. DO NOT disassemble. Refer servicing to qualified service personnel. No user serviceable parts. TO REDUCE THE RISK OF BURNS, ELECTROCUTION, FIRE, OR INJURY TO PERSONS NEVER use near any type of flame or flammable/explosive substances, vapors or atmosphere.
SECTION 1—IMPORTANT SAFETY INSTRUCTIONS NEVER oil or lubricate the compressor coupler or bottle fittings. NEVER use any penetrating oil or lubricant such as WD‐40® or 3‐in‐1 oil®. If connection difficulty is experienced verify that the bottle fitting is being inserted straight down into the compressor coupler. If continued difficulty is experienced contact your supplier for assistance.
SECTION 1—IMPORTANT SAFETY INSTRUCTIONS NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air opening may be blocked. Keep the openings free from lint, hair and the like. If the compressor has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or submersed in water, call a qualified technician for examination and repair. If any leakage of the oxygen cylinder is detected DO NOT attempt to use the cylinder.
SECTION 1—IMPORTANT SAFETY INSTRUCTIONS DO NOT unplug the power cord by pulling on the cord. To unplug, grasp the plug not the cord, otherwise damage may occur. Disposal of Equipment and Accessories Follow local governing ordinance and recycling plans regarding disposal of device components. Radio Frequency Interference This equipment has been tested and found to comply with EMC limits specified by IEC/EN 60601‐1‐2.
SECTION 2—TECHNICAL DESCRIPTION SECTION 2—TECHNICAL DESCRIPTION The HomeFill Oxygen System (IOH200) is an electro‐mechanical prescription system for use in the home. Its intended function and use is to supply pressurized oxygen to fill Invacare proprietary gas cylinders for ambulatory use while the patient receives therapeutic oxygen from an accessory oxygen concentrator.
SECTION 3—INITIAL SETUP SECTION 3—INITIAL SETUP NOTE: Dealers Only ‐ Refer to Dealer Information on page 9 before performing this procedure. Initial Set Up of Compressor Compressor Initial Setup Checklist Before operating the HomeFill Compressor for the first time, complete the following checklist: ❑ Choose a location for the compressor. ❑ Transport the compressor to the desired location. ❑ Connect the power cord to the compressor. ❑ Connect the compressor to the concentrator.
SECTION 3—INITIAL SETUP Transporting the Compressor WARNING NEVER transport a compressor with a cylinder connected to or stored on the compressor, otherwise injury or damage can occur. CAUTION When transporting the compressor, be careful not to bump the compressor or the connectors into obstacles. Otherwise, damage to the compressor may occur. NOTE: For this procedure, refer to FIGURE 3.1. 1.
SECTION 3—INITIAL SETUP Wall Outlet Power Cord Compressor (REAR) Electrical Connection NOTE: This product may come with a polarized plug (one blade is wider than the other). As a safety feature, this plug will fit in a polarized outlet only one way. If the plug does not fit fully in the outlet, reverse the plug. If it still does not fit, contact a qualified electrician. DO NOT attempt to defeat this safety feature. FIGURE 3.
SECTION 3—INITIAL SETUP DETAIL “A” Interconnect Hose Insert End of Interconnect Hose into Inlet/Outlet Fitting CONCENTRATOR TOP VIEW Outlet Fitting Power Cord DETAIL “C” Inlet/Outlet Fitting Metal Tab DETAIL “B” Compressor Inlet Fitting FIGURE 3.
SECTION 4—CHECKLIST AND CYLINDER INSPECTION SECTION 4—CHECKLIST AND CYLINDER INSPECTION Compressor Operation Checklist Each time the HomeFill Compressor is used to fill a cylinder, complete the following checklist: ❑ Ensure the concentrator has been On for at least thirty minutes. Refer to the concentrator Owner’s Manual. ❑ Perform the prefill inspection on the cylinder. ❑ Connect the cylinder to the compressor. ❑ Push the compressor power switch to the On ( ) position.
SECTION 4—CHECKLIST AND CYLINDER INSPECTION HomeFill Cylinder Label Improper filling or abuse of this cylinder, or failure to follow these safety instructions may cause serious bodily injury or death. High pressure oxidizing gas vigorously accelerates combustion. DO NOT alter or modify cylinder or related components. DO NOT use oil or grease. Keep cylinder, cylinder valve, all equipment and connections free of oil and grease to avoid violent ignition resulting in serious bodily injury or death.
SECTION 4—CHECKLIST AND CYLINDER INSPECTION The Hydrostatic Testing Date Aluminum cylinders MUST undergo testing every five years. WARNING DO NOT fill cylinders that have not been tested in the past five years. Otherwise, serious personal injury may result. Contact your dealer for replacement. External Examination 1.
SECTION 5—COMPRESSOR OPERATION SECTION 5—COMPRESSOR OPERATION Connecting/Disconnecting the Cylinder to/from the Compressor WARNING NEVER use tools of any kind to connect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur. Connecting the Cylinder to the Compressor WARNING DO NOT drop oxygen cylinders. Use two hands when handling/transporting oxygen cylinders. Otherwise, injury or damage may occur. DO NOT use a liquid leak detector to test for leaks.
SECTION 5—COMPRESSOR OPERATION 4. Momentarily push DOWN on the outer ring (sleeve) of the connector fillport until GREEN dots are visible to reset the connector (FIGURE 5.1). NOTE: If the outer ring (sleeve) is in the UP position (GREEN dots not visible), the connector fillport will not be able to accept the cylinder fillport. Pushing DOWN momentarily will reset the connector fillport (GREEN dots visible) to accept the cylinder fillport.
SECTION 5—COMPRESSOR OPERATION Cylinder NOTE: Fillport covers not shown for clarity. Cylinder Fillport Power Switch Connector Fillport Compressor Cradle Outer Ring (Sleeve) Compressor Cylinder Fillport Connector Fillport Outer Ring (Sleeve) Cylinder PULL UP FIGURE 5.
SECTION 5—COMPRESSOR OPERATION Disconnecting the Cylinder from the Compressor NOTE: For this procedure, refer to FIGURE 5.3. 1. Press the compressor power switch to the Off (O) position. 2. Grasp the cylinder in the area behind the cylinder fillport. 3. With the other hand, grasp the outer ring (sleeve) of the connector fillport and push DOWN. 4. Lift up on the cylinder to remove from the connector fillport.
SECTION 5—COMPRESSOR OPERATION WARNING The fillport cover on the connector fillport and the cylinder fillport MUST be replaced after filling and whenever not in use. If either of the fillport covers are missing contact your healthcare provider before using. 6. Place the fillport covers onto the connector fillport and cylinder fillport. NOTE: When using the post valve type cylinder, attach an oxygen regulator per manufacturers instructions. 7.
SECTION 5—COMPRESSOR OPERATION Turning the Compressor On NOTE: For this procedure, refer to FIGURE 5.5 on page 29. 1. Make sure the concentrator is On. Refer to the concentrator Owner’s Manual. CAUTION The concentrator flow rate to the patient MUST be set at maximum: 3 L/min or less for Platinum 5, XL, and Perfecto2 and 5 L/min for Platinum 10. Otherwise, the oxygen level to the compressor will be <90% and the compressor will not fill the cylinder. 2.
SECTION 5—COMPRESSOR OPERATION NOTE: The O2 BELOW NORMAL (YELLOW) light may come on if the oxygen level from the concentrator has not yet reached or drops below 90%. The compressor will begin or resume filling when the oxygen level reaches 90%. If light does not go Off within 10 minutes, refer to Troubleshooting on page 31. NOTE: If the ATTENTION (RED) light and audible alarm are on. Refer to Troubleshooting on page 31.
SECTION 6—INDICATOR LIGHTS SECTION 6—INDICATOR LIGHTS INDICATOR LIGHT COLOR COMPRESSOR STATUS None Compressor is Off. O2 BELOW NORMAL YELLOW Compressor is On and warming up (0-3 minutes after power is On). O2 from concentrator is < 90% (any time after three minute warmup). FILLING - GREEN FULL - GREEN ATTENTION - RED (Audible Alarm) HomeFill® Compressor Compressor is On and cylinder is filling. O2 from concentrator is >90% (any time after three minute warmup). Compressor is On.
SECTION 7—TROUBLESHOOTING SECTION 7—TROUBLESHOOTING TROUBLESHOOTING PROBLEM SOLUTION No indicator lights on when compressor turned on. 1. Turn compressor Off. 2. Check that power cord is properly plugged into compressor and wall outlet. 3. Turn the compressor On. 4. Clean air filters. Refer to Cleaning the Compressor Filter on page 33. 5. If performing STEPS 1-4 does not cause any indicator lights to come on, contact your home health care provider or Invacare for service. RED Light ON 1.
SECTION 7—TROUBLESHOOTING TROUBLESHOOTING PROBLEM SOLUTION O2 BELOW NORMAL (YELLOW) light stays On 1. Ensure the concentrator has warmed up for at least 20 minutes.If it has not been 20 minutes, turn the compressor Off until the concentrator warms up. If the YELLOW light still remains on proceed to STEP 2. 2. Inspect the connection between the compressor and the concentrator for damage or pinching. • If damaged, replace. • If pinched, straighten tubing. Turn the compressor Off for at least 10 seconds.
SECTION 8—MAINTENANCE SECTION 8—MAINTENANCE Cleaning the Compressor Filter WARNING Push the power switch to the Off (O) position and unplug the compressor before cleaning. DO NOT operate the compressor without the filter installed. CAUTION DO NOT unplug by pulling on cord. To unplug, grasp the plug not the cord, otherwise damage may occur. NOTE: For this procedure, refer to FIGURE 8.1. NOTE: All compressors have one filter on the back of the cabinet. NOTE: Perform this procedure at least once a week. 1.
NOTES NOTES HomeFill® Compressor 34 Part No 1100873
LIMITED WARRANTY LIMITED WARRANTY NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. This warranty is extended only to original purchaser who purchases this product when new and unused from Invacare Corporation or a dealer.
return authorization date. Do not return products to our factory without prior consent. C.O.D. shipments will be refused; please prepay shipping charges.