Operator’s Manual Platinum™Series XL, 5, 10 Oxygen Concentrators Standard, with SensO ®and HomeFill™II 2 Invacare 5 with SensO2 Model IRC5LXO2 Invacare 5 Model IRC5LX Invacare 10 with SensO2 Model IRC10LXO2 Invacare 10 Model IRC10LX Dealer: This manual MUST be given to the end user. User: Before using this product, read this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.invacare.
WARNING WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT INVACARE TECHNICAL SERVICES BEFORE ATTEMPTING TO USE THIS EQUIPMENT OTHERWISE SERIOUS INJURY OR PROPERTY DAMAGE MAY RESULT. ACCESSORIES There are many different types of humidifiers, oxygen tubing, cannulas and masks that can be used with this device.
TABLE OF CONTENTS ACCESSORIES .................................................. 2 SPECIAL NOTES ............................................... 5 SECTION 1—GENERAL GUIDELINES .................. 6 Maintenance.......................................................................................... 8 Radio Frequency Interference.......................................................... 8 To Reduce The Risk Of Burns, Electrocution, Fire Or Injury To Persons....................................................
TABLE OF CONTENTS SECTION 6—MAINTENANCE .......................... 32 Routine Maintenance........................................................................ 32 SECTION 7—TROUBLESHOOTING GUIDE ......... 35 LIMITED WARRANTY ....................................
SPECIAL NOTES SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words SIGNAL WORD MEANING DANGER Danger indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. WARNING Warning indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
SECTION 1—GENERAL GUIDELINES SECTION 1—GENERAL GUIDELINES In order to ensure the safe installation, assembly and operation of the Platinum concentrator these instructions MUST be followed. CAUTION “Caution: Federal law restricts this device to sale or rental by or on order of a physician, or any other practitioner licensed by the law of the State in which he/she practices to use or order the use of this device.
SECTION 1—GENERAL GUIDELINES WARNING The use of oxygen therapy requires that special care be taken to reduce the risk of fire. Any materials that will burn in air, and some that will not, are easily ignited and burn rapidly in high concentrations of oxygen. For safety concerns, it is necessary that all sources of ignition be kept away from the product and preferably out of the room in which it is being used. NO SMOKING signs should be prominently displayed.
SECTION 1—GENERAL GUIDELINES WARNING NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air opening may be blocked. Keep the openings free from lint, hair and the like. Fill humidifier with water to the level shown by the manufacturer. DO NOT overfill. Keep unit at least three inches away from walls, draperies, furniture, and the like. Invacare recommends that Crush-Proof oxygen tubing be used with this product and not exceed 50 ft. (15.
SECTION 1—GENERAL GUIDELINES WARNING To Reduce The Risk Of Burns, Electrocution, Fire Or Injury To Persons. Avoid using while bathing. If continuous usage is required by the physician’s prescription: The concentrator MUST be located in another room at least seven (7) feet from the bath. DO NOT come in contact with the concentrator while wet. DO NOT place or store product where it can drop into water or other liquid. DO NOT reach for product that has fallen into water. Unplug IMMEDIATELY.
SECTION 1—GENERAL GUIDELINES WARNING Use of some administration accessories or certain humidifiers not specified for use with oxygen concentrator may impair the performance. In certain circumstances oxygen therapy can be hazardous. Invacare recommends that you seek medical advice before using this machine. This equipment is to be used as an oxygen supplement and is not considered life supporting or life sustaining. Avoid creation of any spark near medical oxygen equipment.
SECTION 2—FEATURES SECTION 2—FEATURES Oxygen Purity Indicator Lights (SensO2 model only)/ Fault and Power Indicator Lights (LX model only) Oxygen Outlet Flowmeter Power Switch Circuit Breaker Elapsed Time Meter Cabinet Filter (one of two shown) ACCESSORIES (NOT SHOWN): Remote Humidifier/Flowmeter Accessory IRCRF5 (Platinum 5 and XL Only) REAR VIEW PreciseRX™ Pediatric Humidifier/Flowmeter Accessory - IRCPF16 (Platinum 5 and XL Only) HomeFill II home oxygen compressor IOH200 Power Cord Part No 1118353
SECTION 3—HANDLING SECTION 3—HANDLING The concentrator should ALWAYS be kept in the upright position to prevent cabinet damage while being transported. The shipping container has been designed to assure maximum protection of the concentrator. If the concentrator is to be reshipped by common carrier, additional cartons are available from Invacare. Unpacking NOTE: For this procedure, refer to FIGURE 3.1. 1. Check for any obvious damage to the carton or its contents.
SECTION 4—SPECIFICATIONS SECTION 4—SPECIFICATIONS Electrical Requirements: 120 VAC ± 10% (132 VAC/108 VAC), 60 Hz Rated Current Input: Platinum 5 and XL: 4.3 A Platinum 10: 4.7 A Sound Level: Platinum 5 and XL: 50 dB Average Platinum 10: 58 dB Average Altitude: Platinum 5 and XL - Up to 6,000 ft. (1828 meters) above sea level without degradation of concentration levels. From 6,000 ft. (1828 meters) to 13,129 ft. (4000 meters) below 90% efficiency Platinum 10 - Up to 4,000 ft.
SECTION 4—SPECIFICATIONS 5LXO2 INDICATORS: Label Symbol O2 O2 Purity Indicator Lights SYSTEM FAILURE Below 73% RED - Continuous Audible Alarm Sieve-GARD Compressor Shutdown 73% to 85% YELLOW SYSTEM OKAY Over 85% GREEN 5LX INDICATORS: Label Symbol I/O Status Indicator Lights SYSTEM FAILURE RED SYSTEM OKAY GREEN Flow Range: Platinum 5 and XL - 0.5 to 5 L/min. For flowrates less than 1 L/min., we recommend the use of the Invacare Pediatric Flowmeter IRCPF16 Accessory.
SECTION 4—SPECIFICATIONS Filters: Cabinet (2), Outlet HEPA and Compressor Inlet. Safety System: Current overload or line surge shutdown. High temperature compressor shutdown. High Pressure Alarm w/compressor shutdown. Low Pressure Alarm w/compressor shutdown. Battery Free Power Loss Alarm. SensO2 Oxygen System (SensO2 Model) Low Flow Alarm. Width: 18 3/8 in. ± 3/8 in. (46.7 cm ± 1 cm) Height: 26 3/8 in. ± 3/8 in. (67.0 cm ± 1 cm) Depth: 14 3/8 in. ± 3/8 in. (36.
SECTION 4—SPECIFICATIONS Placement: No closer than three inches from any wall, furniture, draperies, or similar surfaces. Tubing: 7 ft cannula with a maximum 50 ft of Crush-Proof Tubing (DO NOT pinch). Platinum 10: Recommended use up to 50 ft high flow tubing with high flow cannula at all flow rates. Relative Humidity: 20 to 60% Time of Operation: Up to 24 hours per day.
NOTES NOTES Part No 1118353 17 Platinum™Series
NOTES NOTES Platinum™Series 18 Part No 1118353
Please Seal with Tape Before Mailing
REGISTER YOUR PRODUCT! The benefits of registering: 1. 2. 3. 4. 5. Safeguard your investment. Ensure long term maintenance and servicing of your purchase. Receive updates with product information, maintenance tips, and industry news. Invacare can contact you or your provider, if servicing is needed on your product. It will enable Invacare to improve product designs based on your input and needs. Register ONLINE at www.invacare.com - or Complete form and mail.
1. Method of purchase: (check all that apply) ❏ Medicare ❏ Insurance ❏ Medicaid ❏ Other 2. This product was purchased for use by: (check one) ❏ Self ❏ Parent ❏ Spouse ❏ Other 3. Product was purchased for use at: ❏ Home ❏ Facility ❏ Other 4. I purchased an Invacare product because: ❏ Price ❏ Features (list features) 5.
SECTION 5—OPERATING INSTRUCTIONS SECTION 5—OPERATING INSTRUCTIONS Introduction Your oxygen concentrator is intended for individual use in the home. It is an electronically operated device that separates oxygen from room air. It provides high concentration of oxygen directly to you through a nasal cannula. Clinical studies have documented that oxygen concentrators are therapeutically equivalent to other types of oxygen delivery systems. Your provider will show you how to use your oxygen concentrator.
SECTION 5—OPERATING INSTRUCTIONS Recommended Guidelines for Optimum Performance Temperature: 50°F - 95°F (10°C - 35°C) Electrical: No extension cords. Placement: No closer than three inches from the wall, furniture, draperies, or similar surfaces. Tubing and Cannula: 7 ft cannula with a maximum 50 ft of Crush-Proof Tubing (DO NOT pinch). Platinum 10: Recommended use up to 50 ft high flow tubing with high flow cannula at all flow rates. Environment: Smoke and soot-free.
SECTION 5—OPERATING INSTRUCTIONS Connect Humidifier (If So Prescribed) NOTE: For this procedure, refer to FIGURE 5.1 - FIGURE 5.4. NOTE: Fill humidifier with distilled water to the level indicated by the manufacturer. (For Platinum 5, XL - recommended humidifiers are: Invacare 3260, 003-40, or 006-40. For Platinum 10, recommended humidifiers are high flow: 7900-10, 7900-25). WARNING DO NOT overfill humidifier. Replace the humidifier cap and securely tighten.
SECTION 5—OPERATING INSTRUCTIONS 3. Remove the filter access door located on the side of the concentrator. Insert a flathead screwdriver in the plate groove on the top edge of the filter access door and gently pry the filter access door open. Refer to Detail “A” in FIGURE 5.3. 4. The humidifier adapter is next to the inlet filter. Pull up and remove the humidifier bottle adapter. Refer to Detail “B” in FIGURE 5.3. 5.
SECTION 5—OPERATING INSTRUCTIONS 6. Attach oxygen tubing from the humidifier bottle to oxygen outlet connector on the oxygen concentrator (FIGURE 5.4). 7. Attach the cannula/patient supply tubing to the humidifier bottle outlet (FIGURE 5.4). 8. After assembly, ensure that oxygen is flowing through the cannula. Oxygen Outlet Connector Oxygen Tubing Humidifier Outlet FIGURE 5.
SECTION 5—OPERATING INSTRUCTIONS On/Off (I/O) Power Switch FIGURE 5.5 Power Switch Flowrate NOTE: For this procedure, refer to FIGURE 5.6 on page 29. 1. Turn the flowrate knob to the setting prescribed by your physician or therapist. WARNING DO NOT change the L/min. setting on the flowmeter unless a change has been prescribed by your physician or therapist. NOTE: To properly read the flowmeter, locate the prescribed flowrate line on the flowmeter.
SECTION 5—OPERATING INSTRUCTIONS Flow Knob Ball FIGURE 5.6 Flowrate SensO2 Oxygen Purity Indicator (If Your Unit Features the O2 Sensor Feature) This feature monitors the purity of oxygen generated by the oxygen concentrator. If purity falls below factory preset standards, indicator lights on the control panel will illuminate. Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial start warm-up time (approximately 30 min.) while waiting for the O2 purity to reach maximum.
SECTION 5—OPERATING INSTRUCTIONS YELLOW light ( ) - Call supplier IMMEDIATELY. You may continue to use the concentrator unless instructed otherwise by your supplier. Be certain that backup oxygen is nearby. RED light ( )- Total unit shutdown. Switch IMMEDIATELY to backup oxygen supply. Call supplier IMMEDIATELY. GREEN light - with YELLOW light flashing - Call supplier IMMEDIATELY. Oxygen sensor malfunctioning; you may continue to use the concentrator.
SECTION 5—OPERATING INSTRUCTIONS Auto Shut - Down Use Backup RED Call Supplier Normal FIGURE 5.
SECTION 6—MAINTENANCE SECTION 6—MAINTENANCE NOTE: The Invacare concentrators are specifically designed to minimize routine preventive maintenance. In places with high dust or soot levels, maintenance may need to be performed more often. Only authorized or factory-trained personnel should perform preventive maintenance on the concentrator. Power should be disconnected before beginning preventive maintenance on the Invacare concentrator.
SECTION 6—MAINTENANCE CAUTION DO NOT operate the concentrator without the filters installed. Cabinet Filter (located on both sides of concentrator) FIGURE 6.1 Cleaning the Cabinet Filter Cabinet WARNING UNPLUG the concentrator when cleaning. To avoid electrical shock, DO NOT remove cabinet. 1. Clean the cabinet with a mild household cleaner and non-abrasive cloth or sponge. Humidifier NOTE: To clean the oxygen humidifier, follow the instructions provided by the manufacturer.
Platinum™Series 34 *Refer to Preventive Maintenance section of Service Manual. Rebuild Top End of Compressor Replace Exhaust Muffler Clean Heat Exchanger AS REQUIRED Check Power Loss Alarm Check Oxygen Concentration (SensO2 MODELS) Check Compressor Inlet Filter* I RC Check Outlet HEPA Filter* Clean/Replace Cabinet Filter(s) ANNUALLY, DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS Check Oxygen Concentration (UNITS WITHOUT SensO2) MODEL NO.
SECTION 7—TROUBLESHOOTING GUIDE SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: Alarm: Main Power Loss: 1. Insert plug into outlet. SHORT BEEPS, LONG PAUSE 1. Power cord NOT plugged in. 2. Inspect house circuit breakers or fuses. If problem recurs, use a different outlet. Concentrator NOT operating, power switch ON. 2. No power at outlet. 3. Call Service Provider. Beep......Beep...... 3. Tripped circuit breaker. Alarm: CONTINUOUS System Failure: 1.
SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: YELLOW or RED LIGHT Illuminated. Alarm: CONTINUOUS On red light only. * Only applies to 5LXO2 and 10LXO2 SensO2 models. PROBABLE CAUSE: 1. Low oxygen purity.* SOLUTION: 1. Clean or Replace filters. 2. Kinked or blocked tub- 2. Inspect for kinks or blockages. Correct, clean or ing, cannula or humidireplace item. Once corfier.* rected, turn power OFF for 60 seconds and then turn 3. Flowmeter set at 0.5 L/min. for Platinum 5 and power back ON. 1 L/min.
SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: Alarm: RAPID Beep....Beep... Beep...Beep Part No 1118353 PROBABLE CAUSE: SOLUTION: LOW FLOW ALARM 1a. Kinked or blocked tubing, cannula or humidifier. 1a. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power OFF for 60 seconds and then turn power back ON. 1b. Flowmeter set at 0.5 L/min. for Platinum 5 and 1 L/min. for Platinum 10 or less. 37 1b. Reset flowmeter to prescribed flowrate. If prescribed flowrate is below 0.
LIMITED WARRANTY LIMITED WARRANTY NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. This warranty is extended only to the original purchaser who purchases this product when new and unused from Invacare Corporation or a dealer. This warranty is not extended to any other person or entity and it is not transferable or assignable to any subsequent purchaser or owner.
LIMITED WARRANTY For warranty service, please contact Invacare Corporation's Service Department at the toll free number on the back cover. Upon receiving notice of an alleged defect in a product, Invacare Corporation will issue a serialized return authorization.
This product may be manufactured under one or more of the following patents: US 5,906,672 and US 5,917,135. Other patents pending. Invacare Corporation www.invacare.com USA One Invacare Way Elyria, Ohio USA 44036-2125 800-333-6900 Technical Services 800-832-4707 Invacare, Yes, you can., SensO2, and the Medallion Design are registered trademarks of Invacare Corporation. Platinum, HomeFill, Sieve-Gard, Precise RX are trademarks of Invacare Corporation.