Product Manual

TechnicalData
8TechnicalData
8.1ProductParameters
OXYGEN
Oxygen
Measurement
Range:(froma
concentrator)
20.9-96%
Oxygen
Measurement
Accuracy:
±1.5%offullscaleatconstanttemperature
andoptimalflow*
Oxygen
Measurement
Resolution:
0.1%Oxygen
SeeEffectsofLowFlow,page11ofthisoperatingmanual
forinformationontheeffectsoflowflow.
FLOW
FlowMeasurement
Range:
0-10LPM
FlowMeasurement
Accuracy:
±0.2LPM
FlowMeasurement
Resolution:
0.1LPM
PRESSUREPSI
KPA
Pressure
MeasurementRange:
0.5-503.4-344
Pressure
Measurement
Accuracy:
±0.5%±0.5%
Pressure
Measurement
Resolution:
0.10.1upto199,1
from200to344
ResponseTime:
17seconds
Warm-upTime:<1second
Operating
Temperature:
59°F-104°F(15°C-40°C)
Storage
Temperature:
F-140°F(-15°C-60°C)
Pressure:
800-1000mBars
Humidity:
0-95%(non-condensing)
Power
Requirements:
2-AAAlkalinebatteries(2x1.5volts)
BatteryLife:1,100hours(16,500readcycles)
LowBattery
Indication:
“LowBattery”icondisplayedonLCD
Dimensions:
3.16”x5.10”x1.04”(80.3mmx129.5
mmx26.4mm)
Weight:0.4lbs(181grams)
CLASSIFICATIONS
Protectionagainst
electricalshock:
Internallypoweredequipment
Protectionagainst
water:
Modeofoperation:Continuous
Sterilization:See7.1Cleaning,page14.
Flammable
anaestheticmixture:
Notforuseinpresenceofflammable
anaestheticmixtures
Powerspecification:1.8-3.2V 32mW10mA
1176119-A15