Product Manual

1General
1.1Symbols
Signalwordsareusedinthismanualandapplytohazardsorunsafe
practiceswhichcouldresultinpersonalinjuryorpropertydamage.
Seetheinformationbelowfordefinitionsofthesignalwords.
DANGER!
Dangerindicatesanimminentlyhazardoussituation
which,ifnotavoided,willresultindeathorserious
injury.
WARNING!
Warningindicatesapotentiallyhazardoussituation
which,ifnotavoided,couldresultindeathorserious
injury.
CAUTION!
Cautionindicatesapotentiallyhazardoussituation
which,ifnotavoided,mayresultinpropertydamage
orminorinjuryorboth.
Givesusefultips,recommendationsandinformationfor
efficient,trouble-freeuse.
ThisproductcomplieswithDirective93/42/EECconcerning
medicaldevices.
ThelaunchdateofthisproductisstatedintheCE
declarationofconformity.
1.2IntendedUse
WARNING!
TheCheckO
2
Plusisforcheckingoxygen
concentratorsonly.
DONOTusetheCheckO
2
Plusforcontinuous
oxygenmonitoring.
TheCheckO
2
Plusisforuseonlybytrainedpersonnel.
TheCheckO
2
PlusOxygenAnalyzermeasuresoxygenconcentration
andflowusingultrasoundtechnologyandmeasurespressureusinga
piezoresistivesiliconpressuresensor.
TheCheckO
2
PlusOxygenAnalyzerisatoolusedtomeasure
oxygenpurity,flowandpressureattheoutletofanoxygen
concentrator.Itisnotintendedtobeusedbypatientswhoare
prescribedoxygen,norisitintendedtocontinuouslymonitoror
confirmoxygendeliverytoapatient.
TheCheckO
2
PlusOxygenAnalyzerisintendedtobeusedinan
environmentwhereoxygenconcentratorsarebeingservicedor
repaired.ThisincludesHospitals,NursingHomes,ExtendedCare
Facilities,PatientHomes,andRespiratoryDeviceServiceandRepair
Centers.
4