PLATINUM 5 OXYGEN CONCENTRATORS STANDARD, WITH /SENSO2 AND HOMEFILL II Operator's Manual Model IRC5LXO2 IRC5LX

19
TROUBLESHOOTING
1. Internal repairs required.
1. Call Provider for Service.
GREEN LIGHT with
YELLOW Light
Flashing.
Alarm:
SHORT BEEPS,
LONG PAUSE
Concentrator
NOT operating,
power switch ON.
Beep...... Beep......
Main Power Loss:
1. Power cord NOT
plugged In.
2. No power at outlet.
3. Tripped circuit breaker.
1. Insert plug into outlet.
2. Inspect house circuit
breakers or fuses. If
problem recurs, use a
different outlet.
3. Call Provider for
Service.
Alarm:
CONTINUOUS
Power Switch ON,
Concentrator NOT
OPERATING.
Beeeeeeeep....
System Failure:
1. Unit overheating due
to blocked air intake.
2. Insufficient power at
outlet.
3. Internal repairs required.
1a. Remove and clean
cabinet filters.
1b. Move oxygen concentrator
at least three (3) inches
away from walls, draperies
or furniture.
2. Do not use extension cords.
Move to another electrical
outlet or circuit.
3. Call Provider for Service.
YELLOW or RED
LIGHT Illuminated.
*
Only applies to
SensO
2
models.
*1. Low oxygen purity.
*2. Kinked or blocked tubing,
cannula or humidifier.
*3. Flowmeter set at
1/2 L/min. or less.
4. Unit overheating due
to blocked air intake.
5. Insufficient power at
outlet.
6. Internal repairs required.
TROUBLESHOOTING GUIDE
SYMPTOM: PROBABLE CAUSE: SOLUTION:
1. Clean or Replace filters.
2. Inspect for kinks or block-
ages. Correct, clean or
replace item. Once correct-
ed, turn power OFF for 60
seconds and then turn
power back ON.
3. Check flowmeter is set to
1.0 L/min. or more. Refer to
Recommended End-User
Guidelines in the Operating
Instructions of this manual
(page 12).
4.a. Remove and clean cabinet
filters.
4b. Move concentrator at least
3- inches from walls,
draperies, and furniture.
5. Do not use extension
cords. Move to another
electrical outlet or circuit.
6. Call Provider for Service.