User Guide

SECTION 13—CONTROL PANEL
Part No 1148070 53 Perfecto
2
SECTION 13—CONTROL PANEL
Removing Control Panel
DANGER
Topreventelectricalshock,ALWAYSdisconnectconcentratorfromelectricaloutlet
beforeservicing.
NOTE:Forthisprocedure,refertoFIGURE 13.1onpage 54.
ToolsRequired:
•Diagonalcutters
•Flatheadscrewdriver
•¼inchratchet,¼inchsocket,and¼inchextension
Pliers
•Tiewraps
1. Unplugunit.
2. Removecabinet.RefertoCabinetonpage 23.
3. Cutthetwotiewrapsthatsecurethecontrolpaneltothesievebeds(DetailA”).
4. CutthetwotiewrapsthatsecuretheP.E . valveassemblytothecontrolpanel(not
shown).
5. Removethetwomountingscrewsthatsecurethecontrolpaneltothesoundbox
(Detail“B”).
6. Removethe¼inchI.D.tubingfromthetopandbottombarbedfittingsofthe
flowmeter(Detail“B”).
7. Cuttiewrapandremove¼inchI.D.tubingfromthepatientoutletbarbedfitting
behindthecontrolpanel(Detail“B”).
8. RemovethetwomountingscrewsthatsecuretheP.C.boardtothecontrol
panel.
NOTE:BeforeperformingSTEP9,codeallwirestoensurecorrectreinstallation.
9. Removespadeconnectorsfromcircuitbreaker,hourmeterandOn/Off( )switch.
(JumperwireharnessfromOn/Offswitchtocircuitbreakerdoesnothavetobe
removed).
10. Removethehourmeter.RefertoReplacingHourMeter
onpage 64.
11. RemovetheOn/Offswitch.RefertoReplacingOn/OffSwitchonpage 62
12. Removethecircuitbreakerbytwistingthelocknutontheoutsideofthecontrolpanel.
13. Discardexistingcontrolpanel.
14. InstallnewcontrolpanelbyreversingSTEPS312.