SECTION 3—TROUBLESHOOTING
ValueCare Bed Series 16 Part No 1134867
SECTION 3—TROUBLESHOOTING
Troubleshooting Electrical
Troubleshooting Mechanical
SYMPTOM PROBABLE CAUSE SOLUTIONS
Bed spring does not
move.
End of stroke reached. Operate opposite button.
Hi/Lo function does not
work.
Bed not plugged in. Ensure power cord is plugged into power
source.
Pendant not functioning. Ensure Pendant is correctly connected.
Ensure Pendant cable not entangled/pinched.
Cables Ensure cables are not entangled/pinched.
Call the dealer for repair/replacement of
cables and/or motors.
Household fuse or breaker blown. Check household fuse/breaker box.
Motors are not functioning. Call dealer.
Hi/Lo Motor. Ensure cables are not entangled or pinched.
Call dealer.
Thermal Protection Activated. Allow bed unit to sit for 30 minutes and retry.
Bed spring does not stop
moving.
Pendant. Check for depressed/stuck buttons. If button is
stuck, call dealer for repair.
Bed Hi/Lo movement
does not stop.
Check cables, if damaged call dealer.
Bed is producing unusual
sounds, burning odors, or
movement deviations
observed in the controls,
motors, or the limits
switch functions.
Call dealer.
SYMPTOM PROBABLE CAUSE SOLUTIONS
Bed does not stay in
place.
Locking Casters. Lock Casters.
Bed springs do not move.
Pull tube. Check pull tube(s) for binding. Call dealer for
repairs.
Bed ends move in
opposite
directions.
Drive Shaft connected to wrong
gear box shaft.
Call dealer.
Hi/Lo function does not
work.
Drive shaft disconnected. Call dealer.