Product Manual

SECTION 3—TROUBLESHOOTING
Part No 1134867 17 ValueCare Bed Series
WARNING
Read all general guidelines, warnings and cautions before performing any
repair/maintenance on the bed. All repair and maintenance on the bed should be
performed by a qualified technician.
Maintenance Checklist
Invacare recommends the following maintenance and cleaning procedures be conducted
between users and at least once every three years. This form is provided as a guide to help
with documentation.
Mechanical Inspection and Maintenance
Inspect all bed components for damage or excessive wear.
Visually examine all welds for cracks.
Inspect the head and foot springs for bending, warping or damage.
Inspect mounting hardware for bending, damage or excessive wear.
Inspect all bolts and rivets to ensure that they are securely tightened and functioning
properly.
Check sleep surfaces to ensure all links are intact.
Inspect bed height adjustment pin for excessive wear.
Ensure that casters are free of debris.
Check that all labels are present and legible. Replace if necessary.
Electrical Inspection and Maintenance
Check pendant, power and motor cords for chafing, cuts or excessive wear.
Make sure all plugs are fully attached and free of damage.
Check all functions:
Head raises and lowers properly.
Foot raises and lowers properly.
Bed Ends raise and lower properly.
Cleaning and Disinfecting the Bed
Unplug the power cord of electric beds before performing maintenance.
Wipe down with disinfectant. Allow to dry completely.