Product Manual
SECTION 3—TROUBLESHOOTING
Part No 1134867 17 ValueCare Bed Series
WARNING
Read all general guidelines, warnings and cautions before performing any
repair/maintenance on the bed. All repair and maintenance on the bed should be
performed by a qualified technician.
Maintenance Checklist
Invacare recommends the following maintenance and cleaning procedures be conducted
between users and at least once every three years. This form is provided as a guide to help
with documentation.
Mechanical Inspection and Maintenance
❑ Inspect all bed components for damage or excessive wear.
❑ Visually examine all welds for cracks.
❑ Inspect the head and foot springs for bending, warping or damage.
❑ Inspect mounting hardware for bending, damage or excessive wear.
❑ Inspect all bolts and rivets to ensure that they are securely tightened and functioning
properly.
❑ Check sleep surfaces to ensure all links are intact.
❑ Inspect bed height adjustment pin for excessive wear.
❑ Ensure that casters are free of debris.
❑ Check that all labels are present and legible. Replace if necessary.
Electrical Inspection and Maintenance
❑ Check pendant, power and motor cords for chafing, cuts or excessive wear.
❑ Make sure all plugs are fully attached and free of damage.
❑ Check all functions:
• Head raises and lowers properly.
• Foot raises and lowers properly.
• Bed Ends raise and lower properly.
Cleaning and Disinfecting the Bed
❑ Unplug the power cord of electric beds before performing maintenance.
❑ Wipe down with disinfectant. Allow to dry completely.