Product Manual

SECTION 5—MAINTENANCE AND CLEANING
ValueCare Bed Series 28 Part No 1134867
SECTION 5—MAINTENANCE AND
CLEANING
Invacare recommends the following maintenance and cleaning procedures be conducted
between users or as needed. This form may be copied and used as a procedural guide and
documentation.
Date _______________Bed Model #_______________ Bed Serial # _______________
Troubleshooting Electrical
SYMPTOM PROBABLE CAUSE SOLUTIONS
Bed spring does not move. End of stroke reached. Operate opposite button.
Hi/Lo function does not
work.
Bed not plugged in. Ensure power cord is plugged into power
source.
Pendant not functioning. Ensure Pendant is correctly connected. Refer
to Installing the Drive Shaft Assembly
on
page 25.
Ensure Pendant cable not entangled/pinched.
Cables Ensure motor cable(s) are correctly con-
nected. Refer to Installing the Drive Shaft
Assembly on page 25.
Ensure cables are not entangled/pinched.
Call Invacare for repair/replacement of cables
and/or motors.
Hi/Lo Motor. Ensure Hi/Lo motor cable is correctly
connected. Refer to Installing the Drive Shaft
Assembly on page 25.
Ensure cables are not entangled or pinched.
Ensure Drive shaft is correctly installed.
Thermal Protection Activated. Allow bed unit to sit for 30 minutes and retry.
Bed Spring does not stop
moving.
Pendant. Replace pendant. Refer to Installing/Removing
the Pendant on page 27.
Bed Hi/Lo movement
does not stop.
Check cables, if damaged, replace. Refer to
Installing the Drive Shaft Assembly
on page 25.
Bed is producing unusual
sounds, burning odors, or
movement deviations
observed in the controls,
motors, or the limits
switch functions.
Call Invacare.
Full Electric Bed: Hi/Lo
function does not work.
Bed not plugged in. Ensure power cord is plugged into power
source.